1 General description
WARNING!
WARNING!
The devices induce output voltages and currents which conformable to
EN61010-1
are not
dangerous to life. But it is possible that they effect health damages to sensitive persons.
The iseg SHR modules are standalone High Precision HV laboratory SMU – Source Measuring Unit – equipped with the
finest iseg HV generation technology and iCS control system.
The SHR provides up to 4 channels, each with an independent voltage and current control and electronically reversible
polarity. The 6kV channel provides a maximum versatility: with three electronically switchable HV-output modes it can
supply 4mA up to voltages of 2kV, 3mA up to 4kV and 2mA up to 6kV. Alternatively the SHR can be equipped with cost
efficient 2kV/6mA channels.
A high quality 4.3” TFT display shows detailed information and can be controlled by capacitive touch. All comprehensive
features like logging, graphical display and customer specific plugins are also available by the precise jog-wheel and
buttons.
2 Technical Data
SPECIFICATIONS
SHR Standard
SHR High Precision
Polarity
Electronically switchable
Ripple and noise (f > 10 Hz)
< 10 mV
p-p
< 2-3 mV
p-p
Ripple and noise (f > 1 kHz)
< 3 mV
p-p
< 2 mV
p-p
Ripple and noise (10 Hz – 0.1Hz)
< 5-10 mV
p-p
Stability [ΔVout / ΔVin]
2 • 10
-4
• V
mode
1
1
• 10
• 10
-4
-4
•
•
V
V
mode
mode
Stability - [ΔVout / ΔRload]
2 • 10
-4
• V
mode
1
1
• 10
• 10
-4
-4
•
•
V
V
mode
mode
Temperature coefficient
50 ppm/K
30 ppm/K | 10 ppm/K (option TC)
Resolution voltage setting
2 • 10
-6
• V
nom
Resolution current setting
2 • 10
-6
• I
nom
Resolution voltage measurement
2 • 10
-6
• V
nom
1 • 10
-6
• V
nom
Resolution current measurement - full range
2 • 10
-6
• I
nom
1 • 10
-6
• I
nom
Resolution current measurement - 2
nd
range
n/a
50 pA [I
out
< 20 μA]
Accuracy voltage measurement
± (0.01 % • V
out
+0.02 % • V
nom
)
± (0.01 % • V
out
+0.01 % • V
nom
)
Accuracy current measurement - full range
± (0.01 % • I
out
+0.02 % • I
nom
)
± (0.01 % • I
out
+0.01 % • I
nom
)
Accuracy current measurement - 2
nd
range
n/a
± (0.01 % • I
out
+ 4 nA)
Measurement accuracy
- The measurement accuracy is guaranteed in the range 1%
•
• V
mode
< V
out
< V
mode
and for 1 year
Sample rates (SPS)
5, 10, 25, 50, 60, 100, 500
Digital filter averages
1, 16, 64, 256, 512, 1024
Hardware limits
Potentiometer per module [V
max
/ I
max
]; relative to V
nom
/ I
nom
Voltage ramp
1*10
-6
• V
mode
/s up to 0.2 • V
mode
/ s
SHR – Desktop High Voltage Power Supply | Last changed on: 19.01.2018 |
www.iseg-hv.com
5/11