6. Front panel
Figure 2:
Front of CC23
Figure 3:
Front of CC24
System state:
• LED off: Crate is switched off / Standby mode
• LED green: Crate is running without error
• LED red: Error occurred (Crate Status bit Sum Error is set)
LED STATUS
LED STDBY
Standby mode:
• LED blue: Crate is connected to mains,
remote controllable, WiFi and Ethernet connections are active
Pushbutton
POWER ON
Switches the crate on or off by pressing the button for one second
ON: All modules will be supplied with power by crate and can be controlled
The fan speed is regulated according to the maximum temperature within the crate
OFF: All modules will be switched off. The fans run at minimal speed.
LED HV ON
HIGH VOLTAGE is active
LED yellow: At least one channel in crate generates high voltage or is measuring an
output voltage of more than 60 Volt
CAN1
RJ-45 CAN – connector for slave controller connection
CAN2
RJ-45 CAN – connector for slave controller connection
ETH (CC24 only)
RJ-45 Ethernet network connector (10/100/1000MBit)
LED SLAVE 0-3
(CC23 only)
Indicate the slave position in the yellow or green CAN line
USB (CC24 only)
USB 2.0 connector, for external WiFi, camera, firmware upgrade, flash memory
SL 1
Safety-Loop-Connector 1
SL 2
Safety-Loop-Connector 2
CONTROL
D-SUB-9 connector with configurable digital I/O
The voltage supply of the crates backplane and the crate connected modules is switched on by the
POWER ON
pushbutton or by
remote control.
Crate Controller CC24/CC23 | Last change on: 14.04.2020 |
www.iseg-hv.com