70375-EN Rev B
11
Pattern Selection Screen
A
Displays selected pattern.
B
Right control knob selects
Radius
in microns (Triple Arc, Circle only). The radius is the distance
from the origin to the inner edge of nearest treatment spot. There will be a different minimum and
maximum Radius range based on the selected treatment spot diameter; for example, a pattern
with a 100-micron spot will have a minimum Radius of 500 microns. Examples:
C
Right control knob selects
Spacing
between spots (Grid, Triple Arc, Circle only). The spacing is
the distance between the inner edges of a pair of spots. Spacing is displayed as increments of
spot size diameters, and is adjustable from 0.0 to 3.0 in 0.25 spot size increments; for example, a
pattern with a 100-micron spot with 1.00 spacing will have a 100-micron spacing between spots.
Examples:
D
Confirms pattern scanning selection and returns to Treatment Screen.
E
Middle control knob selects pattern type: Grid 2x2, Grid 3x3, Grid 4x4, Grid 5x5, Grid 6x6,
Grid 7x7, Triple Arc, Circle.
F
Displays total number of laser spots for selected pattern.
B
A
C
D
E
H
G
F