PASCAL
®
Synthesis Ophthalmic Scanning Laser Systems
Page 75 of 141
88292-EN, Rev D
Single Spot
Primary pattern
Secondary pattern
15ms EpM
Single Spot
Not available
General Use
•
PRP, retina tears, retina detachments, iridotomy, trabeculoplasty, LIO
Pattern
Spot Diameter (μm)
EPM %
Single Spot
100 μm
10 to 95
200 μm
The Single Spot pattern can be used with the slit lamp or an optional laser indirect
ophthalmoscope (LIO). Spot diameter and repetition rate settings are adjustable.
NOTE
The Power and Exposure settings cannot be adjusted for 15ms EpM Single Spot pattern. The
Power and Exposure settings keep the same as the previous selected EpM pattern.
Endpoint Management Features
Endpoint Management On/
Landmark Off
Single endpoint spots
(orange-xx %)