
PASCAL
®
Synthesis Ophthalmic Scanning Laser Systems
Page 113 of 141
88292-EN, Rev D
No aiming beam is present when in READY mode and/or no laser
treatment light is delivered when the footswitch is depressed and/or
the beams are of poor quality.
Probable Cause:
Laser is in STANDBY mode, not READY mode.
Suggestion:
Select READY mode on control panel.
Probable Cause:
Footswitch is not connected.
Suggestion:
Connect footswitch.
Probable Cause:
Aiming beam is on low intensity setting.
Suggestion:
Adjust aiming beam intensity on control panel.
Probable Cause:
After five minutes of non-use, system goes to STANDBY.
Suggestion:
Switch mode from STANDBY to READY.
Probable Cause:
Remote interlock has been activated and has disabled
system.
Suggestion:
Ensure the action that activated remote interlock has ceased
and proceed.
Probable Cause:
Footswitch and/or footswitch cable damaged.
Suggestion:
Inspect for damage.
Probable Cause:
Internal system error.
Suggestion:
Contact service.