NOTE
These ratings only apply to laser exposure greater than 200 mm (7.9 in) away from the laser
exit aperture of the SLA.
The type of eye protection recommended for the physician, the patient, and/or treatment
room personnel within the NHZ depends on the planned procedure and the equipment
required to perform that procedure.
An eye safety filter is provided with the SLA and is required for safe use. Laser safety
eyewear is not required by the physician who views the procedure through the slit lamp
eyepieces. All other personnel within the NHZ must wear laser safety eyewear with the
recommended optical density.
Along with providing the appropriate safety eyewear, the following steps should be taken to
secure the controlled area:
1. Treatment should be conducted in a dedicated, enclosed room.
2. A warning sign should be placed on the outside of the treatment room door when the
laser is in use. The sign is intended to alert personnel before they enter the controlled
area.
3. The treatment room door should be kept closed during treatment.
Electrical Hazards
WARNING: DANGEROUS VOLTAGE
To avoid the risk of electric shock, this equipment must only be connected to a supply mains
with protective earth. Hospital Grade Cord grounding reliability can only be achieved when
equipment is connected to an equivalent receptacle marked “Hospital Grade” or “Hospital
only”.
To avoid the risk of electric shock, do not touch any external connector and the patient
simultaneously.
Do not use power cables other than the power cable provided with the system. Do not use
extension cables with the system.
Disconnect the laser system from the electrical outlet when inspecting the fuses.
Never open the laser console protective covers. Opening the covers will expose you to high
voltage components, the laser resonator and possible laser radiation. Only certified
personnel shall work inside the console.
The area around the laser and footswitch shall be kept dry. Do not operate the laser if any of
the cords are faulty or frayed. The laser should undergo routine inspection and maintenance
per the Iridex Corporation manufacturer's recommendation and institutional standards.