background image

4 | Proper DERS Bolus Programming

*

a. To program a bolus, press 

Bolus Softkey A/B

 or

    press 

Menu Button.     

e. 

PRESS

 the 

TIME

  button,  program  time      

    using  

Numeric Keypad.

f. Press

 ENTER 

to Start Bolus.

11

Programming a Bolus

d. Program dose using 

Numeric Keypad

c. Press 

Dose

 Softkey.

b. From Menu screen, press 

Softkey

 by 

    

Bolus A 

or 

Bolus B.

 Enter the infusion time of 6 seconds (0.1 minutes) to 99  

   minutes.

12

KVO (Keep Vein Open)

e. Press 

Start/Stop A or B Button 

to resume infusion.

5 | Resuming Infusion During KVO

a. Press 

Start/Stop A or

 

B

 (depending on channel in use) to 

    return to running screen.

b. Press the

 VTBI 

Softkey.

c. Program new VTBI

 

using

 Numeric Keypad

d. Press 

ENTER

     Ready to Start

 will  appear at bottom of the screen.

Note:

 

Primary infusion will resume following completion of bolus.

*

Bolus option ONLY available if enabled by facility pharmacy.

Summary of Contents for MRIDIUM 3860+

Page 1: ...MRidium MRidium 3860 3860 MRI Infusion Pump MRI Infusion Pump Reference Guide Reference Guide www IRadimed com www IRadimed com...

Page 2: ...issues during infusions 1 1 2 2 3 3 1 1 Selecting the Proper IV Set Post Prime Post Prime a Select the 1056 IV Set when infusing from a bag or bottle fluid source 18mL b Select the 1057 IV Set when i...

Page 3: ...pump then push to insert as shown The clamp will click when secured DO NOT Push the Black Flow Clamp which triggers a Check Door Alarm and cause possible Freeflow Initial Setup c Secure tubing inside...

Page 4: ...NTER again Program entries with Numeric Keypad such as Dose ENTER Weight kg ENTER VTBI ENTER Programming Channel A a Press the softkey next to New Patient Drug Library A or Press the softkey next to N...

Page 5: ...BI ENTER Programming Channel B 7 Programming Basic Infusions 3 1 Rate Mode Channel A d Press the Start Stop A Button to start infusion as shown in 2 1 a Press the softkey by New Patient Rate Mode on M...

Page 6: ...TBI Softkey Program the VTBI using the Numeric Keypad Press the ENTER key Ready to Start will appear at bottom of the screen 9 Titrating Basic Infusions 3 3 Rate Mode Channel A or B During Infusion In...

Page 7: ...Bolus B Enter the infusion time of 6 seconds 0 1 minutes to 99 minutes 12 KVO Keep Vein Open e Press Start Stop A or B Button to resume infusion 5 Resuming Infusion During KVO a Press Start Stop A or...

Page 8: ...ical Department for repair Alarm Messages 6 2 PT Occlusion Alarm Alarm will sound if occlusion between the pump and patient occurs Check between the pump and patient to assure that All Clamps have bee...

Page 9: ...e Pump from use and contact your facility s Biomedical Department and or IRadimed Technical Support at 1 866 677 8022 Notes _____________________________________________________ ______________________...

Page 10: ...struction Information subject to change IRadimed Corporation 2020 All rights reserved LB026 A 1025 Willa Springs Drive Winter Springs FL 32708 407 677 8022 Always be aware of the Flow Clamp Position R...