background image

5

Before Programming Channels A or B

2.1 | Selecting Infusion Type: Drug Libraries A or B

a. Press the 

UP

 or 

DOWN 

arrow keys to scroll through listed   

    medications. Press 

ENTER

 when the proper medication 

    is highlighted. 

d. To start infusion, press the

 Start/Stop A Button.

b. Primary is highlighted, press 

ENTER

 again.

    Program entries with 

Numeric Keypad

 such as:

                                                                 -  Dose; 

ENTER

                                                      - Weight (kg); 

ENTER

                                                                    - VTBI; 

ENTER

Programming Channel A

a. Press the softkey next to 

New Patient (Drug Library A)

or

Press the softkey next to 

New Patient (Drug Library B)

b. Press the softkey next to 

Dose Rate Calc. A

or

Press the softkey next to 

Dose Rate Calc. B

6

2.2 | Selecting Medication: Drug Library A

Use Library A for Channel A, and Library B for Channel B

c. 

Push Start A

 will appear when programmed 

    correctly.

Startup Screen

Menu Screen

Summary of Contents for MRIDIUM 3860+

Page 1: ...MRidium MRidium 3860 3860 MRI Infusion Pump MRI Infusion Pump Reference Guide Reference Guide www IRadimed com www IRadimed com...

Page 2: ...issues during infusions 1 1 2 2 3 3 1 1 Selecting the Proper IV Set Post Prime Post Prime a Select the 1056 IV Set when infusing from a bag or bottle fluid source 18mL b Select the 1057 IV Set when i...

Page 3: ...pump then push to insert as shown The clamp will click when secured DO NOT Push the Black Flow Clamp which triggers a Check Door Alarm and cause possible Freeflow Initial Setup c Secure tubing inside...

Page 4: ...NTER again Program entries with Numeric Keypad such as Dose ENTER Weight kg ENTER VTBI ENTER Programming Channel A a Press the softkey next to New Patient Drug Library A or Press the softkey next to N...

Page 5: ...BI ENTER Programming Channel B 7 Programming Basic Infusions 3 1 Rate Mode Channel A d Press the Start Stop A Button to start infusion as shown in 2 1 a Press the softkey by New Patient Rate Mode on M...

Page 6: ...TBI Softkey Program the VTBI using the Numeric Keypad Press the ENTER key Ready to Start will appear at bottom of the screen 9 Titrating Basic Infusions 3 3 Rate Mode Channel A or B During Infusion In...

Page 7: ...Bolus B Enter the infusion time of 6 seconds 0 1 minutes to 99 minutes 12 KVO Keep Vein Open e Press Start Stop A or B Button to resume infusion 5 Resuming Infusion During KVO a Press Start Stop A or...

Page 8: ...ical Department for repair Alarm Messages 6 2 PT Occlusion Alarm Alarm will sound if occlusion between the pump and patient occurs Check between the pump and patient to assure that All Clamps have bee...

Page 9: ...e Pump from use and contact your facility s Biomedical Department and or IRadimed Technical Support at 1 866 677 8022 Notes _____________________________________________________ ______________________...

Page 10: ...struction Information subject to change IRadimed Corporation 2020 All rights reserved LB026 A 1025 Willa Springs Drive Winter Springs FL 32708 407 677 8022 Always be aware of the Flow Clamp Position R...

Reviews: