1.3 | Power
ON
the Pump
Press the
( I )
button as shown to power on the pump.
Initial Setup
1.4 | Open the Door
“Push Down” on the
Purple Button
, “Pull Out” on the
Handle.
Purple
Button
Handle
1.5 | Loading the IV Set
a. Insert the round
Disk
on the IV set into the “
Disk Slot
” on top of the pump.
b. Tilt up and insert the top of the
Black Flow
Clamp
into the pump, then push
to insert as shown. The clamp will click when secured.
-
DO NOT
Push the
Black Flow Clamp
which triggers a
Check Door Alarm
and cause possible
Freeflow.
Initial Setup
c. Secure tubing inside
Bubble Detector
Door Pin
d. Firmly
CLOSE THE DOOR
Ensure
Door
Handle
is
FULLY Latched,
Open
All
Clamps
,
Vent Caps.
No fluid flow should be in the drip chamber of the IV set or bubbles
in a syringe. If fluid flow is present, clamp the IV tubing and repeat
steps 1.2 on page 2.
Note:
Contact your biomed department if the door pin does not look as shown.