pg. ii_DaqCal
918494
User Calibration Guide, 457-0931 rev 1.0
Equipment
Refer to the appropriate block below, depending on whether you will be calibrating
through a 37-pin, 68-pin, or 100-pin connector
.
P1
37-PIN
If calibration signals will be passing through a 37-pin P1 connector you will need
:
Required
:
6.5-digit, digital multi-meter
1
adjustable voltage calibrator
2
ambient temperature meter
In addition, the following are recommended
:
DBK11A, passive screw-terminal board
CA-37, 37 conductor ribbon cable
P4 100-PIN
If calibration signals will be passing through a 100-pin P4 connector you will need
:
Required
:
6.5-digit, digital multi-meter
1
adjustable voltage calibrator
2
ambient temperature meter
In addition, the following are recommended
:
DBK11A, passive screw-terminal board*
DBK200, P4-to-DB37 adapter card*
CA-195 100-conductor ribbon cable
*Another DBK200 Series device that includes a P1 and P4 connector may be substituted.
See page iv.
P5 SCSI 68-PIN
If calibration signals will be passing through a 68-pin SCSI (P5) connector you will need
:
Required
:
6.5-digit, digital multi-meter
1
adjustable voltage calibrator
2
ambient temperature meter
In addition, the following are recommended
:
TB-100, passive screw-terminal board
CA-G56 (3 ft. shielded cable) **
** One of the following cables may be substituted: CA-G55, a 3 ft. unshielded cable; or
CA-G56-6, a 6 ft. shielded cable.
Certain steps in the calibration process require that you apply a signal to the device via a specific pin, or
take a measurement reading from a specific pin. To make the task easier we recommend that you use a
passive screw-terminal board, such as the DBK11A, TB-100, or possibly a DBK200 Series device.
See the
Setup
section beginning on page iii.
1
Agilent 34401A or better DMM.
2
Traceability is through the DMM. The calibrator is only used as a quiet and stable voltage source.
Summary of Contents for DaqBoard/3000 Series
Page 2: ...ii...
Page 6: ...vi DaqBoard 3000 Series User s Manual...
Page 16: ...IG 8 DaqBoard Installation Guide 918494 1033 0940 rev 8 0...
Page 38: ...2 8 Connections Pinouts 918494 DaqBoard 3000 Series User s Manual...
Page 42: ...4 2 Calibration 918494 DaqBoard 3000 Series User s Manual...
Page 82: ...6 12 Setpoint Configuration for Output Control 908794 DaqBoard 3000 Series User s Manual...
Page 104: ...DBK215 pg A 14 948894 Appendix A...
Page 106: ...B 2 Hardware Analog Level Trigger 90794 Appendix B...