Replacement of Non-rechargeable Batteries
BATTERY REPLACEMENT:
NEVER REPLACE PRIMARY ALKALINE BATTERY CELLS
WHILE IN A POTENTIALLY EXPLOSIVE OR HAZARDOUS LOCATION. USE ONLY
INDUSTRIAL BY DURACELL ID1500 BATTERIES.
Fitting batteries or connecting the Battery Pack with the wrong polarity may
result in damage to the instrument.
WEEE REGULATIONS
Dispose of used batteries in accordance with all local and national safety and
environmental requirements.
Inlet Probe Assembly
All parts of the probe assembly can be replaced if they become damaged or contaminated with use.
1 Probe
–
Tiger (880207)
4 O-ring (5/OV-02)
2 Filter Housing Cap (912221)
5 Filter Clamp (912220)
3 Probe Seal (880202)
PTFE Filter Disc
It is essential that the TIGER XTS is always used with a supplied 0.5 micron PTFE
Filter Disc fitted to the front of the instrument. Without the filter, particles of
debris and dust can be drawn into the MiniPID 2 Sensor inhibiting the function
of the instrument. These filters are consumable and should be changed after
every 100 hours of use. The frequency of replacement should be increased for
dusty or moisture laden environments. PTFE Filter Discs are available from your
distributor or at
PTFE Filter Disc changing should be conducted in a suitably clean environment, with clean hands and
equipment to avoid contamination of the new PTFE Filter Disc.