TIGER Select Instrument User Manual V3.2
Unrivalled Gas Detect
ion.
Page 11 of 73
ionscience.com
2.
Product Overview
GASES THAT ARE CROSS SENSITIVE TO BENZENE WILL RESULT IN SIGNIFICANT
ERRORS IN READING.
The TIGER Select is a hand-held gas detector for the rapid, accurate detection of benzene and Total
Aromatic Compounds (TACs).
Utilising the high output ION Science 10.0 eV detection system, a reading for TACs is seen
immediately on start-up. Should aromatics be detected, an ION Science Benzene Pre-filter Tube can
be easily attached to ensure fast detection and selective measurement of benzene. Benzene Pre-
filter Tubes absorb many VOC gases, but benzene will pass through.
Benzene
Various
VOC gases
Direction of gas flow
The number of gases a PID can detect directly relates to the maximum photon energy (eV) of the PID
Lamp being used. As TIGER Select uses a 10.0 eV light source, many gases associated with benzene
are ignored due to their higher eV.
Throughout the measurement process, TIGER Select continues to display real-time data, ensuring
the final reading represents the full value of actual benzene present. Benzene concentrations are
displayed down to parts-per-billion (ppb) levels, giving you accurate, reliable data you can count on.
The TIGER Select is capable of providing 15-minute short-term exposure limits (STELs) and 8-hour
time-weighted averages (TWAs) for total aromatic compounds (TACs). In addition, the ION Science
MiniPID 2 Sensor incorporates both humidity resistant technology and anti-contamination design for
extended operation in difficult working environments.
The TIGER Select benzene and total aromatic compound detector can also be used in standard
operational mode without the use of a Benzene Pre-filter Tube to deliver active indications of
volatile organic compounds (VOCs), including benzene, at concentrations as low as 1 ppb benzene
equivalent.
The TIGER Select has been designed and certified Intrinsically Safe.