Batteries
Replacement of Non-rechargeable Batteries in Battery Pack
WARNING
BATTERY REPLACEMENT
:
Never replace primary Alkaline battery cells while in a potentially
explosive or hazardous location. Use only Duracell Procell Alkaline
batteries MN1500.
a.
Ensure TIGER
LT
is switched Off.
b.
Remove the AA Battery Pack (
A3
) as in (b) and (c) on page 34;
c.
Remove the screws retaining the battery
cover and lift it off. This reveals a set of 3 x
AA cells contained in a removable battery
holder;
d.
Remove the battery holder;
e
Replace the exhausted batteries; Use
Duracell PROCELL Alkaline batteries
MN1500
f.
Check all batteries have the correct polarity
before replacing the battery holder;
g.
Assemble the battery holder into the AA
Battery Pack (
A3
), refit the battery cover and
fix in place using the securing screws.
Replace the Instrument Boot (
8
) ensuring that
it is a snug fit around the front of the
Instrument Body (
A1
) and the lights are not
obscured.
CAUTION:
Fitting batteries or connecting the battery pack with the wrong polarity may result in permanent
damage to the instrument!
CAUTION:
Non-rechargeable batteries must not be replaced in the field. Batteries should be loaded into the
battery holder in a safe environment only. Only the assembled AA Battery Pack (A3) may be
changed in the field.
CAUTION:
Do not make any connections to the USB port of this instrument whilst in a hazardous area.
Note:
When loading batteries check for correct battery polarity before connecting them.
Note:
Dispose of used batteries in accordance with all local and national safety and environmental