TIGER
LT
MANUAL
Ion Science Ltd
Page 31 of 41
Diagnostics
Basic faults or diagnostics are presented as symbols. Should a fault occur most can be corrected by
pressing
Enter
or
Esc
to clear the fault message. All fault conditions cause the TIGER
LT
to alarm.
Pump failure
Pump blocked or pump failure
The flow of gas through the instrument has fallen below 100cc/minute. Check the
probe and filter for signs of blockage. Water or dirt in the probe, a bent probe, dirty
filter on the inlet or blockage of the exhaust (finger over hole on the back?) can all
cause low flow. If the blockage can be removed, press
Esc
to clear the alarm. If the
fault persists send the instrument to your distributor for service.
Battery dead
Battery low or battery failure
The TIGER
LT
will shut down when the battery level falls below 2%. Recharge the
battery as instructed in the ‘Batteries’ section of this manual ensuring that all
connections are sound and the indicator lights on the charger are in order. If the
battery fails to charge, fit another battery pack if available. If using alkaline batteries,
replace them. If the fault persists send the instrument and charger to your distributor
for service.
Lamp out
Lamp failure
The PID lamp has failed to strike (illuminate); this may occur at switch on or during
use. Switch the TIGER
LT
off and replace the lamp. See the ‘Maintenance’ section.
Memory full
Memory cannot receive more data
The data log memory is full. This will only happen if the Log Full box is set to ‘alarm’
on the TIGER
LT
PC configuration screen. Press the
Esc
key to continue, but the
TIGER
LT
will no longer continue to log data. Select ‘recycle’ in TIGER PC
LT
and the
TIGER
LT
will overwrite the oldest data and no alarm will be raised.
System error
Total system failure
The instrument’s firmware is corrupted. In the unlikely event of this message
appearing, contact Ion Science Ltd or your nearest authorised service centre.