15
Rev 05
Geometry
Coil Selection
3IB-MRIDevic
-channel Combination
2345
-connection
d
Multi Coil
no
Homogeneity correction none
FOV (mm)
400.00
RFOV (%)
100.00
Foldover suppression
no
Matrix scan
256
-reconstruction
256
Scan percentage (%)
50.00
Stacks
3
-current
A
-type
parallel
-slices
3
-slice thickness (mm)
10.00
-slice gap
user defined
- - gap (mm)
10.00
-slice orientation
sagittal
-foldover direction
AP
Slice scan order
default
Stack display order
no
PlanAlign
no
REST slabs
0
Patient position
head first
-orientation
supine
Contrast
Scan mode
M2D
-technique
FFE
Contrast enhancement
T1
Fast Imaging mode
TFE
-shot mode
multi-shot
TFE factor
42
-shot interval
shortest
-profile order
linear
Echoes
1
-partial echo
no
TE
shortest
Flip angle (deg)
20.00
TR
user defined
-(ms)
15.00
Half Scan
no
Water fat shift
maximum
Shim
no
Run a SURVEY scan using the parameters listed below.
TIP: Save this SURVEY scan for future QA scans.
An example of the SURVEY images is on page 15.
SPIR
no
TFE prepulse
invert
-slice selection
no
-delay
shortest
MTC
no
SAR mode
default
Gradient mode
regular
Motion
Cardiac synchronisation
no
Respiratory compensation
no
Flow compensation
no
NSA
1
Dyn/ang
Angio
no
Quantitative flow
no
Manual start
no
Dynamic study
no
Postproc
Preparation phases
auto
MIP/MPR
no
Images:
O:M no no no
Autoview image
M
Reference tissue
White matter
Preset window contrast
soft
Reconstruction mode
immediate
Save raw data
no
Push to workstation
no
Hardcopy protocol
no
Ringing filter
no
Silent mode
no
Offc/ang
no
Stacks
3
-current
A
Stack Offc.AP (P=+mm)
-20.00
RL (L=+mm)
0.00
FH (H=+mm)
0.00
Ang. AP (deg)
0.00
RL (deg)
-0.00
FH (deg)
-0.00
Summary of Contents for Invivo 500082
Page 27: ...27 Rev05...