InMode Without Vasculaze Operator Manual
Section 9: Fractora Treatment
Information
57 of 96
Establish and confirm treatment parameters: if the desired end-point of
erythema and edema – in a tip-shaped pattern – has not been achieved within
10-15 minutes, increase the RF energy. If the response is excessive, decrease the
parameters.
Tolerance of patient to the treatment: low energy levels of 10-25 may be
tolerated after topical anesthesia application. Energy levels of 25-40 may require
nerve block or tumescent and high energy levels of 40-62 may call for tumescent
or IV sedation.
2.2
Treatment Recommendations
1
.
Apply the necessary means of anesthesia. If topical, make sure that it is
cleaned off the face before treatment and the skin is dried with alcohol 70%.
2
.
Ensure that skin is clean and dried with alcohol 70%.
3
.
Take an alcohol cleaned and dry tip and connect it to the Handpiece in the
groove.
4
.
Connect Handpiece to the system.
5
.
Follow Device Start-Up Procedure from System Operation section.
6
.
Set treatment parameters. Typical treatment parameters are shown in the
table below:
Clinical
Indications
Tip
Treatment
Area
RF
Energy
Level
Number
Sessions
Weeks
between
Sessions
Down
Time
Days
Superficial
- texture,
tone, very
fine lines
60
Soft tissue
10-25
3-5
1-4
0-1
Bony area
5-20
Moderate -
texture,
tone, fine
lines
60
Soft tissue
25-30
2-3
4-5
2-3
Bony area
20-30
Deep -
texture,
tone, fine
lines, fine
vessels,
wrinkles,
acne scars
60
Soft tissue
30
1-2
5-6
7-10
Bony area
52
-
3
0