![Invacare SOLO2 TPO100B Service Manual Download Page 20](http://html1.mh-extra.com/html/invacare/solo2-tpo100b/solo2-tpo100b_service-manual_2082624020.webp)
SECTION 5—TROUBLESHOOTING AND MAINTENANCE
SOLO
2
™
20
Part No 1164898
Error Codes
.
ERROR CODE
DESCRIPTION
COMMENT
Common Alarms
256
Warning - Battery less than 25%
Charge battery, see Owner’s Manual
512
Warning - Battery less than 15%
Charge battery, see Owner’s Manual
768
Warning - No Breath Found
Check Canula
1024
Warning - Battery less than 10%
Charge battery, see Owner’s Manual
1280
Warning - Breath Rate High
Reduce Activity
1536
Alarm - Unit Temp HI/LO
Check ambient temp and blockage of air
inlet/outlet
1792
Warning - Start Temp HI/LO
Check ambient temp and blockage of air
inlet/outlet
2048
Alarm - Battery Temp HI/LO
Check ambient temp and blockage of air
inlet/outlet
2049
Alarm - Battery Temp HI/LO
Check ambient temp and blockage of air
inlet/outlet
2304
Warning - Low O2 Purity
Check ambient temp and blockage of air
inlet/outlet
2560
Alarm - Low O2 Purity
Check ambient temp and blockage of air
inlet/outlet
3584
Alarm - Stuck Button
Remove power, restart unit
3840
Alarm - Low Flow Output
Check Canula and patient outlet filter
4352
Warning - EXT PWR Low
See Owner’s Manual
4608
Alarm - EXT PWR Low
Find New Power
4096
Alarm - Fan is not moving
Operating Retry - Call Provider
4097
Alarm - High pressure
Operating Retry - Call Provider
4098
Alarm - Low pressure
Operating Retry - Call Provider
4099
Alarm - Max shift time exceed
Operating Retry - Call Provider
ERROR CODE
DESCRIPTION
COMMENTS
Compressor Alarms
3328
Alarm - Compressor lock
3328
3329
Alarm - Starting speed check
3329
3330
Alarm - Power on but compressor
shutdown
3330
3331
Alarm - Power not on but compressor told
to start
3331