SECTION 2—COMPONENT REPLACEMENT
Part No 1164898
11
SOLO
2
™
Replacing the Inlet Filter
NOTE:
For
this
procedure,
refer
to
FIGURE 2.2.
NOTE:
The
inlet
filter
should
be
replaced
according
to
the
Transportable
Oxygen
Concentrator
Preventive
Maintenance
Record
on
page 23.
1. Position
Transportable
Oxygen
Concentrator
flat
on
the
backside
of
unit
to
locate
the
inlet
filter
access
door
(DETAIL
“A”).
2. Using
a
#2
Phillips®
screw
driver,
remove
the
retaining
screw
securing
the
inlet
filter
access
door
to
unit
(DETAIL
“B”).
3. Remove
inlet
filter
access
door
from
unit.
4. Locate
existing
inlet
filter
and
remove
(DETAIL
“C”).
5. Install
new
replacement
inlet
filter.
6. Reinstall
inlet
filter
access
door
by
inserting
tabs
into
the
slots
located
on
the
unit.
7. Reinstall
retaining
screw
to
secure
the
inlet
filter
access
door.
NOTE:
Use
only
Invacare
replacement
Part
Number
1157083
(Inlet
Filter)
on
the
Transportable
Oxygen
Concentrator.
FIGURE 2.2
Replacing the Inlet Filter
Filter Access
Door
DETAIL “A”
DETAIL “B”
Inlet Filter
Screw
DETAIL “C”