Invacare® Pronto® Air Personal Transporter with MyBody Seating
1.
Remove the battery box. Refer to 8.2.11 Removing/Installing the
Battery Box From/Into the Personal Transporter, page 47.
2.
Remove the hex head screw locknut
A
, located on the underside
of the personal transporter base frame
B
, that secures the
footboard
C
to the footboard tube
D
.
3.
Remove the footboard from the footboard tube.
If adjusting the footboard depth is necessary, align one
of three mounting holes in the footboard assembly with
the mounting hole in the footboard tube.
4.
Reverse STEPS 1-2 to install/adjust the depth of the footboard
assembly.
6.8.2
Adjusting the Footboard Height
1.
Remove the socket head screw
A
, three washers
B
, spacer
C
and locknut
D
that secures the footboard
E
to the footboard
support
F
.
2.
Remove the footboard from the footboard support.
3.
Position the footboard onto the footboard support so that the
mounting holes in the personal transporter frame align with the
desired mounting holes in the footboard support.
4.
Using the socket heat screw, three washers, spacer and locknut
secure the footboard to the footboard support.
6.8.3
Adjusting the Footboard Angle
1.
Loosen the jam nut
A
and set screw
B
located on the rear of
the footboard
C
.
2.
Adjust the set screw in or out to obtain the desired footboard
angle.
3.
Thread the jam nut until it is flush with the footboard bracket.
4.
Securely tighten the jam nut against the footboard to secure
the set screw in place.
34