Invacare PreciseRx IRCPF16 Owner'S Manual Download Page 6

Invacare PreciseR

x

™ Pediatric Flowmeter

6

Part No 2001105

Setting The Flowrate On The Pediatric Flowmeter

1.

Set the pediatric flowmeter to the setting prescribed by your physician.

2.

Center the ball of the Pediatric Flowmeter with the line of the prescribed flowrate. This will ensure 

an accurate flowrate. 

3.

Place cannula on the patient.

FIGURE 3   

Setting The Flowrate On The Pediatric Flowmeter

5 Care and Maintenance

5.1

Cleaning

The unit can be cleaned with warm soapy water. Rinse with a soft damp cloth and dry thoroughly.
The humidifier bottle MUST be cleaned after each use.
Humidifier bottle and tubing MUST be changed after each patient.

5.2

Service and Preventative Maintenance

The PreciseR

x

™  Pediatric Flowmeter is specifically designed to minimize routine preventive 

maintenance.
If damage to the unit is suspected, or if for any reason flow accuracy is in question, Invacare recommends 

returning the unit to Invacare for calibration by qualified personnel. 

For this procedure, refer to FIGURE 3.

WARNING

DO NOT select a different setting unless your physician has prescribed a change in the 
flowrate of the PreciseR

X

 

Pediatric Flowmeter.

CAUTION

DO NOT allow water or other substances to enter flowmeter openings. DO NOT 

submerge flow stand, flowmeter or tubing.

Flowrate on the standard concentrator flowmeter will drop below the 2.0 L/min. setting 

after the Invacare PreciseR

Pediatric Flowmeter is installed and operating.

CAUTION

DO NOT allow water or other substances to enter flowmeter openings. DO NOT 

submerge flow stand, flowmeter or tubing.

Flowmeter

Ball

Summary of Contents for PreciseRx IRCPF16

Page 1: ...in this manual and apply to hazards or unsafe practices which could result in personal injury or property damage See the information below for definitions of the signal words WARNING Warning indicates...

Page 2: ...n is being delivered NO SMOKING signs should be prominently displayed Textiles and other materials that normally would not burn are easily ignited and burn with great intensity in oxygen enriched air...

Page 3: ...ed flowmeter in a dry area 2 DO NOT place other objects on top of the repackaged device DESCRIPTION QUANTITY Flowmeter Stand 1 Crush Proof Tubing CP 25 ft Max 7 5 m 1 Crush Proof Tubing 7 ft 2 1 m tot...

Page 4: ...cannula on patient only after all tubing connections have been made Possible water feedback through the tubing could occur if this sequence is not followed When adjusting the Pediatric Flowmeter the f...

Page 5: ...2 1 m 7 ft max oxygen tubing from the pediatric flowmeter output fitting to the inlet of the humidifier bottle using an appropriate adapter if applicable 5 Connect 7 6 m 25 ft max oxygen tubing to the...

Page 6: ...e The PreciseRx Pediatric Flowmeter is specifically designed to minimize routine preventive maintenance If damage to the unit is suspected or if for any reason flow accuracy is in question Invacare re...

Page 7: ...ion It shall then be the responsibility of the purchaser to return the entire unit or remove at purchaser s cost the defective component part s identified pack the component part s in a manner to avoi...

Page 8: ...8300 800 668 5324 Invacare Corporation 2101 E Lake Mary Blvd Sanford FL 32773 800 832 4707 2010 Invacare Corporation All rights reserved Republication duplication or modification in whole or in part i...

Reviews: