Invacare PreciseRx IRCPF16 Owner'S Manual Download Page 2

Invacare PreciseR

x

™ Pediatric Flowmeter

2

Part No 2001105

2 Safety

WARNING

DO NOT USE THIS PRODUCT OR ANY AVAILABLE OPTIONAL EQUIPMENT 

WITHOUT FIRST COMPLETELY READING AND UNDERSTANDING THESE 

INSTRUCTIONS AND ANY ADDITIONAL INSTRUCTIONAL MATERIAL SUCH 

AS OWNER’S MANUALS, SERVICE MANUALS OR INSTRUCTION SHEETS 

SUPPLIED WITH THIS PRODUCT OR OPTIONAL EQUIPMENT. IF YOU ARE 

UNABLE TO UNDERSTAND THE WARNINGS, CAUTIONS OR INSTRUCTIONS, 

CONTACT A HEALTHCARE PROFESSIONAL, DEALER OR TECHNICAL 

PERSONNEL BEFORE ATTEMPTING TO USE THIS EQUIPMENT - OTHERWISE, 

INJURY OR DAMAGE MAY OCCUR.

ACCESSORIES WARNING

Invacare products are specifically designed and manufactured for use in conjunction with 

Invacare accessories. Accessories designed by other manufacturers have not been 

tested by Invacare and are not recommended for use with Invacare products.

WARNING

DO NOT SMOKE while using this device. Keep all matches, lit cigarettes or other 

sources of ignition out of the room in which this product is located and away from 

where oxygen is being delivered.
NO SMOKING signs should be prominently displayed. Textiles and other materials that 

normally would not burn are easily ignited and burn with great intensity in oxygen 

enriched air. Failure to observe this warning can result in severe fire, property damage 

and cause physical injury or death.
A spontaneous and violent ignition may occur if oil, grease or greasy substances come in 

contact with oxygen under pressure. Example of items to avoid are hand creams or 

lotions, cooking oils, suntan oils or sunblocks and similar oily products. If you use these 

types of products, wash your hands prior to using this product. These substances MUST 

be kept away from the oxygen concentrator, tubing and connections, and all other 

oxygen equipment. DO NOT use any lubricants unless recommended by Invacare.
The PreciseR

x

 Pediatric Flowmeter is designed for an input pressure of 34.5 kPa (5 psi) 

only. The flowmeter is applicable only with Invacare concentrators with 34.4 kPa (5 psi) 

output pressure. 
Keep out of reach of children.
DO NOT disassemble. Refer servicing to qualified service personnel.

Check ALL parts for shipping damage. If shipping damage is noted, DO NOT use. 

Contact carrier/dealer for further instructions.

Summary of Contents for PreciseRx IRCPF16

Page 1: ...in this manual and apply to hazards or unsafe practices which could result in personal injury or property damage See the information below for definitions of the signal words WARNING Warning indicates...

Page 2: ...n is being delivered NO SMOKING signs should be prominently displayed Textiles and other materials that normally would not burn are easily ignited and burn with great intensity in oxygen enriched air...

Page 3: ...ed flowmeter in a dry area 2 DO NOT place other objects on top of the repackaged device DESCRIPTION QUANTITY Flowmeter Stand 1 Crush Proof Tubing CP 25 ft Max 7 5 m 1 Crush Proof Tubing 7 ft 2 1 m tot...

Page 4: ...cannula on patient only after all tubing connections have been made Possible water feedback through the tubing could occur if this sequence is not followed When adjusting the Pediatric Flowmeter the f...

Page 5: ...2 1 m 7 ft max oxygen tubing from the pediatric flowmeter output fitting to the inlet of the humidifier bottle using an appropriate adapter if applicable 5 Connect 7 6 m 25 ft max oxygen tubing to the...

Page 6: ...e The PreciseRx Pediatric Flowmeter is specifically designed to minimize routine preventive maintenance If damage to the unit is suspected or if for any reason flow accuracy is in question Invacare re...

Page 7: ...ion It shall then be the responsibility of the purchaser to return the entire unit or remove at purchaser s cost the defective component part s identified pack the component part s in a manner to avoi...

Page 8: ...8300 800 668 5324 Invacare Corporation 2101 E Lake Mary Blvd Sanford FL 32773 800 832 4707 2010 Invacare Corporation All rights reserved Republication duplication or modification in whole or in part i...

Reviews: