Invacare PreciseRx IRCPF16 Owner'S Manual Download Page 5

Part No 2001105

5

Invacare PreciseR

x

™ Pediatric Flowmeter

Set Up With A Humidifier Bottle If Prescribed

1.

Connect a 2.1 m (7 ft) max oxygen tubing from the output fitting of the concentrator to the input 

fitting on the pediatric flowmeter.

2.

Follow the directions supplied by the humidifier bottle manufacturer for instructions on filling and 

connecting to oxygen tubing.

3.

After filling the humidifier bottle with water, place it in the humidifier compartment on the Invacare 

concentrator (see concentrator owner’s manual for mounting the bottle).

4.

Connect 2.1 m (7 ft) max oxygen tubing from the pediatric flowmeter output fitting to the inlet of 

the humidifier bottle (using an appropriate adapter if applicable).

5.

Connect 7.6 m (25 ft) max oxygen tubing to the outlet of the humidifier bottle (not shown).

6.

Attach the opposite end of the oxygen tubing to a nasal cannula using an appropriate oxygen tubing 

connector (not shown).

FIGURE 2   

Typical Setup - Set Up With A Humidifier Bottle If Prescribed

For this procedure, refer to FIGURE 2.

CAUTION

DO NOT coil crush resistant oxygen tubing on COLD floor or condensation will form 

inside the tubing.

2.1 m (7 ft) Oxygen 

Tubing

Output Fitting

Concentrator

Pediatric 

Flowmeter

Output Fitting

Input Fitting

2.1 m (7 ft) Oxygen Tubing

Inlet of 

Humidifier Bottle

Humidifier 

Bottle

Summary of Contents for PreciseRx IRCPF16

Page 1: ...in this manual and apply to hazards or unsafe practices which could result in personal injury or property damage See the information below for definitions of the signal words WARNING Warning indicates...

Page 2: ...n is being delivered NO SMOKING signs should be prominently displayed Textiles and other materials that normally would not burn are easily ignited and burn with great intensity in oxygen enriched air...

Page 3: ...ed flowmeter in a dry area 2 DO NOT place other objects on top of the repackaged device DESCRIPTION QUANTITY Flowmeter Stand 1 Crush Proof Tubing CP 25 ft Max 7 5 m 1 Crush Proof Tubing 7 ft 2 1 m tot...

Page 4: ...cannula on patient only after all tubing connections have been made Possible water feedback through the tubing could occur if this sequence is not followed When adjusting the Pediatric Flowmeter the f...

Page 5: ...2 1 m 7 ft max oxygen tubing from the pediatric flowmeter output fitting to the inlet of the humidifier bottle using an appropriate adapter if applicable 5 Connect 7 6 m 25 ft max oxygen tubing to the...

Page 6: ...e The PreciseRx Pediatric Flowmeter is specifically designed to minimize routine preventive maintenance If damage to the unit is suspected or if for any reason flow accuracy is in question Invacare re...

Page 7: ...ion It shall then be the responsibility of the purchaser to return the entire unit or remove at purchaser s cost the defective component part s identified pack the component part s in a manner to avoi...

Page 8: ...8300 800 668 5324 Invacare Corporation 2101 E Lake Mary Blvd Sanford FL 32773 800 832 4707 2010 Invacare Corporation All rights reserved Republication duplication or modification in whole or in part i...

Reviews: