Invacare® Platinum™ Mobile
Delivered Oxygen Pulse Volumes:
•
The nominal pulse volumes published in the following table is in milliliters at STPD (standard temperature and pressure dry) conditions and
apply over the rated ranges for ambient temperature, humidity, and atmospheric pressure.
•
Maximum variation from nominal: +/- 15%
4 ft to 25 ft Cannula/Oxygen Tubing Lengths
Breaths Per Minute
15
20
25
30
35
40
Pulse Setting = P1
14.7
11.0
8.8
7.3
6.3
5.5
Pulse Setting = P2
29.3
22.0
17.6
14.7
12.6
11.0
Pulse Setting = P3
44.0
33.0
26.4
22.0
18.9
16.5
Pulse Setting = P4
58.7
44.0
35.2
29.3
25.1
22.0
8.3
Regulatory Listing
ETL certified
complying with:
EN/IEC 60601-1; Ed: 3.1
EN/IEC 60601-1-2; Ed: 4
AAMI ES60601–1 (United States)
CSA 22.2 No. 60601–1 (Canada)
ISO 80601–2–69
ISO 80601–2–67
IEC 60601–1–6
IEC 60601–1–8
IEC 60601–1–11
RTCA DO 160G
64
1194969-A-00