Invacare® Platinum™ Mobile
LCD DISPLAY TEXT:
DESCRIPTION:
SOLUTIONS:
No Breath Detected
Check Cannula
INDICATORS:
Single audible beep every 30 seconds
YELLOW alarm indicator illuminated
Oxygen concentrator has not detected a
breath over a 15 second period.
1.
Verify cannula is connected, not kinked,
properly positioned and user is breathing
through his/her nose.
NOTES:
Alarm will turn off when a breath is detected.
This alarm does not apply during warm up.
If no breath is detected for two minutes, the alarm will escalate to a shutdown alarm. Refer to Shutdown Alarms in Troubleshooting.
The first occurrence of this alarm condition will cancel the audio off (mute) feature. Audio off (mute) can be reactivated. If reactivated, the
alarm audio off (mute) feature will remain activated even if this alarm condition occurs again.
LCD DISPLAY TEXT:
DESCRIPTION:
SOLUTIONS:
Breath Rate High
Reduce Activity
INDICATORS:
Single audible beep every 30 seconds
YELLOW alarm indicator illuminated
The user breath rate has exceeded the capacity
of the concentrator for more than 15 seconds.
1.
The user should immediately reduce
his/her activity level to get his/her breath
rate to slow down.
NOTES:
A sufficient reduction of the user’s breath rate will automatically turn off the alarm.
The first occurrence of this alarm condition will cancel the audio off (mute) feature. Audio off (mute) can be reactivated. If reactivated, the
alarm audio off (mute) feature will remain activated even if this alarm condition occurs again.
48
1194969-A-00