SECTION 20—4-WAY VALVE/MANIFOLD ASSEMBLY
Part No 1164934
65
Perfecto
2
™
V
SECTION 20— 4-WAY VALVE/
MANIFOLD ASSEMBLY
DANGER
To
prevent
electrical
shock,
ALWAYS
disconnect
concentrator
from
electrical
outlet
before
servicing.
CAUTION
This
is
a
maintenance
free
valve.
Opening
the
valve
will
void
any
and
all
warranties
applicable
to
the
valve.
Replacing 4-Way Valve and/or Manifold Assembly
NOTE:
For
this
procedure,
refer
to
FIGURE 20.1
on
page 67.
1. Unplug
unit.
2. Remove
cabinet.
Refer
to
Removing
Cabinet
on
page 23.
3. Cut
the
one
tie
wrap
that
secures
the
manifold
exhaust
tube
to
the
“F”
tube.
4. Remove
the
manifold
exhaust
tube
from
the
“F”
tube.
5. Cut
the
two
tie
wraps
that
secure
the
“F”
tube
to
the
4
way
valve/manifold
assembly.
6. Remove
the
“F”
tube
from
the
4
way
valve/manifold
assembly.
7. Using
a
hose
clamp
tool,
remove
the
clamps
that
secure
the
vinyl
tubes
to
the
4
‐
way
valve/manifold
assembly.
8. Remove
the
vinyl
tubes
from
both
sides
of
the
4
way
valve/manifold
assembly.
9. Note
the
orientation
of
the
4
‐
way
valve/manifold
assembly
before
removing
for
correct
re
‐
installation.
10. Lift
4
‐
way
valve/manifold
assembly
up
and
out
of
the
base
of
the
concentrator.
11. Disconnect
spade
connectors
located
on
the
side
of
the
4
‐
way
valve
assembly
(Not
shown).
12. Remove
the
four
mounting
screws
that
secure
the
4
‐
way
valve/manifold
assembly
together
(Detail
“A”).
13. Replace
the
necessary
component(s).
(i.e.
‐
4
‐
way
valve,
manifold
or
both).
14.
Loosely
install
mounting
screws
through
4
‐
way
valve
and
into
manifold
assembly.
NOTE:
Ensure
the
manifold
gasket
is
in
place
on
4
‐
way
valve
before
installing.