SECTION 6—PREVENTIVE MAINTENANCE
Perfecto
2
™
V
30
Part No 1164934
ON EACH INSPECTION
MODEL NO.
I RC
SERIAL NO.
PREVENTIVE MAINTENANCE RECORD
Record Date of Service
Record Elapsed Hours On Hour Meter
Clean Cabinet Filter(s)
Check Prescribed L/min. Flow Rate
EVERY
26,280 HOURS OR 3 YEARS, WHICHEVER COMES FIRST
Check Oxygen Concentration
Clean/Replace Cabinet Filter(s)
Check Outlet HEPA Filter*
Check Compressor Inlet Filter*
Check Power Loss Alarm
AS REQUIRED
Clean Heat Exchanger
Replace Exhaust Muffler
Rebuild Top End of Compressor
*NOTE:
Refer to Preventive Maintenance section of Service Manual.
NOTE: 4,380 hours are equivalent to usage 24 hours per day, 7 days per week, for 6 months.
DURING PREVENTIVE MAINTENANCE SCHEDULE, OR BETWEEN PATIENTS
FIGURE 6.7
Preventative Maintenance Record