SECTION 24—ALARM TEST
Part No 1154245
89
Perfecto
2
™
4. Slowly adjust flow beyond maximum rated flow until you can achieve a concentration
level greater than 75% but less than 84%. Within 30 minutes, the YELLOW
panel
indicator light will illuminate and the unit will continue to run.
5. Clamp off the 1/8-inch oxygen sensor tubing between the oxygen sensor and the
product tank regulator.
WARNING
Use care not to cut oxygen sensor tubing when clamping the oxygen sensor tubing.
6. Within 30 minutes, the Oxygen Sensor Alarm should activate and the RED panel light
indicator will illuminate. The compressor will shut down with the alarm.
FIGURE 24.3
Oxygen Sensor
SENSO
2
ALARM THRESHOLD
LABEL
SYMBOL
STATUS
INDICATOR LIGHTS
SYSTEM OKAY
O
2
over 85%
GREEN Indicator Light
O
2
Between 73% to 85%
YELLOW Indicator light
A. YELLOW Solid
B. YELLOW Flashing Sensor Failure
Call a qualified technician.
SYSTEM FAILURE
O
2
Below 73%
RED Indicator light
Continuous Audible Alarm Sieve-GARD™Compressor
Shutdown
Call a qualified technician
1/8-inch I.D. Oxygen Sensor Tubing
(Clamp Off Flow)
O
2