SECTION 5—INDICATOR LIGHTS
32
ATTENTION - RED
(Audible Alarm)
Compressor is On.
Cylinder NOT filling
Causes:
Full cylinder attached
when unit turned On -
replace with empty
cylinder
or
Internal failure -
contact supplier.
Compressor does not
start filling or stops
filling
Unit MUST be turned
Off and On before filling
can occur.
O
2
flowing to user if
concentrator is On.
INDICATOR LIGHT
COLOUR
COMPRESSOR
STATUS
EFFECT