10 TROUBLESHOOTING
Part No 1153410
67
Invacare ®CS Series Beds
Actuator not working.
Faulty hand control pendant.
Pendant cable is connected, power
is on, lock-out is off and hand
control pendant does not function.
Disconnect power supply then
replace hand pendant with
functioning unit (from another bed,
if available). Reconnect power and
test.
Contact Invacare Continuing Care Service
Department 1-800-668-2337.
Bed ends to not fit
properly.
Bent brackets.
Visually ensure the brackets are
parallel to the frame.
Carefully straighten minor deformations. Call Service
Department to replace significantly bent brackets,
1-800-668-2337.
EFFECT
POSSIBLE CAUSE
VERIFICATION
CORRECTIVE ACTION