5.9 FIRST BREATH - Air supply capacity for S breathing
valve
1. Connect Test adapter MPE-P2 (K) to the Test connection (3).
2. Connect the black tube of the Test adapter MPE-P2 (K) to the Adapter breathing
valve #2 (N).
3. Connect the Adapter breathing valve #2 (N) to the breathing valve.
4. Close the positive pressure lever on the breathing valve.
5. Slowly open the cylinder valve.
6. Hold the Test adapter MPE-P2 (K) in your hand and carefully manually activate (ON)
the positive pressure lever on the breathing valve.
7. Record the value on the Low pressure gauge (2) and if the value rises above 10 mbar
then close the positive pressure lever.
8. Close the cylinder valve.
9. Release the remaining pressure.
10. Disconnect Test adapter MPE-P2 (K) and Adapter breathing valve #2 (N) from the
breathing valve.
If a complete SCBA set is to be tested – go to section “6. Leak test of complete
apparatus with mask”.