User manual 

 

32749D01 

 

Summary of Contents for Spirotest 100

Page 1: ...ENGLISH Spirotest 100 Model 2016 User manual 32749D01...

Page 2: ...cted by copyright All rights are reserved Interspiro and Divator are registered trademarks belonging to Interspiro This publication may not be copied photocopied reproduced translated or converted to...

Page 3: ...eathing valve 15 4 5 Opening pressure of exhalation valve in Spiromatic breathing valve 15 4 6 Static pressure of Spiromatic breathing valve 15 4 7 Automatic activation of positive pressure for Spirom...

Page 4: ...Spirotest 100 Model 2016 4...

Page 5: ...st 100 Model 2016 respectively of its components with these operating instructions as well as with the purposes of use mentioned therein or as confirmed by letter In any case Interspiro recommends hav...

Page 6: ...r the operation of the Spirotest 100 Model 2016 further test accessories are required depending on the equipment to be tested It is absolutely crucial to read the following instructions carefully and...

Page 7: ...tion button 3 Test connection 12 Test head deflation button 4 Test head 13 Positive pressure button 5 Relief valve 14 Negative pressure button 6 High pressure connection 15 Exhalation button 7 Selecto...

Page 8: ...6 290 704 I Test adapter BV P Divator 336 290 706 J Rubber plug 97 381 01 K Test adapter MPE P2 336 290 659 L Adapter for breathing resistance S mask 98 597 01 M Adapter breathing valve 1 98 595 01 N...

Page 9: ...sure gauge 2 7 Disconnect Sealing cap RD 40 A from the Test connection 3 8 Adjust the Low pressure gauge 2 to zero if necessary 9 Low pressure test complete 2 2 High pressure leakage test 1 Connect HP...

Page 10: ...pen the Relief valve 5 until the test value is reached on the High pressure gauge 1 b Compare the value on the High pressure gauge 1 to the value shown on the pressure gauge of the SCBA set c Record t...

Page 11: ...the Medium pressure gauge 9 shows 7 5 bar 0 5 bar d Check the dynamic medium pressure by repeating step 3 above NOTE When adjusting it is important always to go from a higher pressure 9 bar to a lower...

Page 12: ...otocol 3 Push Release button 17 to release remaining pressure 4 Disconnect the Adapter hose ST C95 E from the Medium pressure connection 18 5 Disconnect the HP adapter from the High pressure connectio...

Page 13: ...of approximately 100 bar is reached on the High pressure gauge 1 then close the cylinder valve 6 Slowly open the Relief valve 5 7 When air is audibly relieved from the breathing valve record the value...

Page 14: ...approved test value 8 Disconnect the Adapter hose ST P49 G from the Test connection 3 9 Disconnect the breathing valve from the Adapter hose ST P49 G Troubleshooting Should a not allowed leak occur t...

Page 15: ...SCBA set cylinder valve and manifold 2 Connect the hose from HP adapter to the High pressure connection 6 3 Connect the medium pressure hose from SCBA set to the breathing valve on the Test connection...

Page 16: ...e Test connection 3 4 8 Air supply capacity for Spiromatic breathing valve 1 Connect Test adapter MPE P2 K to the Test connection 3 2 Connect the black tube of the Test adapter MPE P2 K to the inhalat...

Page 17: ...he Low pressure gauge 2 by pressing the Vacuum pump button 10 and the Negative pressure button 14 at the same time 7 Start the stop watch and after 15 seconds record the value on the Low pressure gaug...

Page 18: ...sure gauge 2 by pressing the Vacuum pump button 10 and the Negative pressure button 14 at the same time 5 Start the stop watch and after 15 seconds record the value on the Low pressure gauge 2 and com...

Page 19: ...p watch and after 15 seconds record the value on the Low pressure gauge 2 in the test protocol and compare with the approved test value 7 Close the cylinder valve 8 Release the remaining pressure 9 Di...

Page 20: ...positive pressure lever on the breathing valve 5 Slowly open the cylinder valve 6 Release the pressure slightly by pressing the Positive pressure button 13 for 2 seconds 7 Start the stop watch and aft...

Page 21: ...ing valve 5 Slowly open the cylinder valve 6 Hold the Test adapter MPE P2 K in your hand and carefully manually activate ON the positive pressure lever on the breathing valve 7 Record the value on the...

Page 22: ...1 to stabilize 5 Close the cylinder valve and record the value on the High pressure gauge 1 6 Start the stop watch and after 1 minute check the value on the High pressure gauge 1 for any decrease in p...

Page 23: ...Spirotest 100 Model 2016 23 7 Technical Data Width 400 mm Depth 380 mm Height 340 mm Weight 16 5 kg excl test adapters and accessories Mains supply 110 220 V AC 50 60 Hz...