1. INTRODUCTION

The ILT770 is a hand held UV meter designed to measure both irradiance in Watts per square
centimeter (W/cm²) and dose/exposure in Joules per square centimeter (J/cm²). The system 
comes comfingured with one UV sensors and is calibrated as a set (paired meter/sensor).  
Sensors cannot be swapped from one meter to the other as this will impact the accuracy of 
the readings.

The ILT770 includes, a storage case, mini USB cable, light meter, UV sensor and NIST 
traceable calibration with a certificate.

2. BUTTON FEATURES

The ILT770 meter has a membrane panel with four buttons that allows numerous processes.

  

1 POWER / UNITS

Press and Hold: Turn device on or off

Press:  Toggle units of readouts, W/cm²or JW/cm²

Note: The ILT770 is auto ranging and will automatically scale to the correct level. A prefix will 
be applied to values below 1W/cm2 ( ie mW/cm2, µW/cm2, mJ/cm2 and µJ/cm2)

  2  ZERO  / HOLD

  

Press and Hold:  ZERO: Takes a reading and subtracts that value from all future readings.

2A. Dark ZERO: Cover sensor, wait 2-3 seconds, press zero. This is used to apply a 
proper zero for the electronics in the working environment.

2B. Ambient ZERO: Place the sensor in the measurement location with the test source 
off or covered. .Wait 2-3 seconds for the system to take a measurement of the ambient 
light levels. Apply a zero to subtract the ambient from future readings. Turn the lamp on
and measure only the light supplied from the test lamp.

 Press:  Freezes/Holds the readings on the display

 3 RECORD

Press to save/store the results on the display in the internal memory of the ILT770 Meter.

Record saves the following data:

ILT770 UV METER

Page number 3 or 9 

Summary of Contents for ILT770

Page 1: ...ILT770 UV Light Meter Operation Manual ILT770 UV METER Page number 1 or 9 10 Technology Drive Peabody MA 01960 Ph 978 818 6180 Fax 978 818 6180 Web www intl lighttech com...

Page 2: ...duction 3 2 Button Features 3 3 Taking a Measurement 5 3A Taking a measurement first time users 5 4 Graphing 6 5 Battery 7 6 Sensor 7 7 USB Ports 7 8 Calibration 7 9 Warranty 8 10 Software 8 ILT770 UV...

Page 3: ...or JW cm Note The ILT770 is auto ranging and will automatically scale to the correct level A prefix will be applied to values below 1W cm2 ie mW cm2 W cm2 mJ cm2 and J cm2 2 ZERO HOLD Press and Hold...

Page 4: ...Press 3 Save now DATA Shows the data analysis from when Graph was first pressed including Min Max and Average NOTE you cannot return to this screen to save the results Press 3 Save now EXIT Returns th...

Page 5: ...ense any changes to the light level and will provide a readout on the display updating every second Press button 3 to save the readings as needed Press and hold button 1 to shut off your device or wai...

Page 6: ...4 once to enter graph mode and take your measurements Press again to stop measuring and review graph Press a third time to review the analysis and finally press a fourth time to return to the main dis...

Page 7: ...should be at the same distance from the source as the product being irradiated The Sensor is not water proof and should not be submerged Liquids should be used very sparingly when cleaning the sensor...

Page 8: ...y in connection with said equipment 10 SOFTWARE The ILT770 software package allows the user to apply firmware updates configure meta data to help differentiate readings for owners of multiple systems...

Page 9: ...ILT770 UV METER Page number 9 or 9...

Reviews: