background image

Units of readout in W/cm²

 

or JW/cm²  as shown on display.

Date/Time Stamp

Numerical Values

Graphical Analysis Values (when in Graph mode)

     Min reading 

     Max readings 

     Average  

The ILT770 has the ability to apply meta data to the saved results. See Applying meta data in 
chapter 10 software,  on page 8.

  4 GRAPH MODE

The Graph button has four functions  (Start Graph, Stop/Review Graph, Data Analysis and 
Exit) that cycle 

forward

 each time button 4 is pressed. It is not possible to return to a previous

screen. 

GRAPH:  Changes the display to graphical mode and start plotting the results and temporarily
storing the readings for final data analysis. 

STOP:      Stops graphing and holds the display showing the last 60 results. 

(NOTE: you cannot return to this screen, to save the results, Press 3, Save now)

DATA:      Shows the data analysis from when Graph was first pressed, including Min, Max, 
and Average.

(NOTE: you cannot return to this screen, to save the results, Press 3, Save now)

EXIT:      Returns the display to numerical /non graphing mode and clears the previous graph 
data from the display. 

 

ERASE ALL SAVED DATA

Press and hold both buttons 3 & 4 at the same time to delete all saved reading from the 
internal memory of the ILT770.

3. TAKING A MEASUREMENT

The ILT770 sensor is calibrated with the sensor aperture over filled with light perpendicular to 
the light source. The calibration reference plane is the front surface of the housing. When 
taking an irradiance or dose measurement the same conditions should be applied. The light 
source to be measured must fully cover the input optics opening with light.  The sensor should
be placed so that the front surface is at the treatment location and perpendicular to the 
source.

ILT770 UV METER

Page number 4 or 9 

Summary of Contents for ILT770

Page 1: ...ILT770 UV Light Meter Operation Manual ILT770 UV METER Page number 1 or 9 10 Technology Drive Peabody MA 01960 Ph 978 818 6180 Fax 978 818 6180 Web www intl lighttech com...

Page 2: ...duction 3 2 Button Features 3 3 Taking a Measurement 5 3A Taking a measurement first time users 5 4 Graphing 6 5 Battery 7 6 Sensor 7 7 USB Ports 7 8 Calibration 7 9 Warranty 8 10 Software 8 ILT770 UV...

Page 3: ...or JW cm Note The ILT770 is auto ranging and will automatically scale to the correct level A prefix will be applied to values below 1W cm2 ie mW cm2 W cm2 mJ cm2 and J cm2 2 ZERO HOLD Press and Hold...

Page 4: ...Press 3 Save now DATA Shows the data analysis from when Graph was first pressed including Min Max and Average NOTE you cannot return to this screen to save the results Press 3 Save now EXIT Returns th...

Page 5: ...ense any changes to the light level and will provide a readout on the display updating every second Press button 3 to save the readings as needed Press and hold button 1 to shut off your device or wai...

Page 6: ...4 once to enter graph mode and take your measurements Press again to stop measuring and review graph Press a third time to review the analysis and finally press a fourth time to return to the main dis...

Page 7: ...should be at the same distance from the source as the product being irradiated The Sensor is not water proof and should not be submerged Liquids should be used very sparingly when cleaning the sensor...

Page 8: ...y in connection with said equipment 10 SOFTWARE The ILT770 software package allows the user to apply firmware updates configure meta data to help differentiate readings for owners of multiple systems...

Page 9: ...ILT770 UV METER Page number 9 or 9...

Reviews: