10. Trigger: TTL, for synchronization of measurement. Currently not available.

The ILT5000 does not include a trigger measurement configuration.  

11. Dark and Zero settings:

Each ILT5000 is pre-programmed with the factory dark (“zero”) setting with no sensor attached in a low noise 
environment.  Because ambient environments differ, ILT software allows the user to program and select the 
“user dark”.  Using the user dark feature will allow the customer to subtract small amounts of current 
generated from their environment.  Because the ILT5000 is a highly sensitive device, User Dark settings may 
be required for improved accuracy in low light / current level measurements.

12. ILT5000 initial set up:

12A. Software Installation:

 

   The first step is to download the software d on the ILT website: 

http://www.intl-lighttech.com/support/software

 

 

 (see the “DataLight III Software Suite….” link). 

 The download will include a Windows Installer file with a .msi extension.  When the msi file is executed, you 
are brought through a simple set of “Next”, “Next”, “Finish” steps. At the completion, a DataLight folder will 
be placed on the desk top with the available app.s located inside the folder.

NOTE:

If you are restricted from or have trouble downloading the ILT software, please email: 

ilsales@intl-

lighttech.com

 and we will gladly email you a copy.

12 B. Device driver installation:

(in most computers the FTDI Drivers are not already present).  After 

installation of DataLight III, device drivers are saved in C, Program Files (x86), International Light 
Technologies, Datalight III, FTDI Driver.  Double click on the CDM executable file to install the drivers.

13. Verifications
13A. USB Com Port Verification: 

After the drivers are installed, turn on the ILT5000, plug the USB 

cable into the computer and ILT5000 ports.  Wait a few minutes to allow comp port assignment. 

When the device is properly loaded, a “USB Serial Port” will show up in Device Manager under “Ports (COM &
LPT)” as seen in the image below.

ILT5000 Research Radiometer Manual, Rev B, 4/17                                                                                     6 of 13

Summary of Contents for ILT5000

Page 1: ......

Page 2: ...as with warning Page 5 8 Response time Page 5 9 Flash N A Page 5 10 Trigger Page 6 11 Dark and Zero settings Page 6 12 Set up and Installations Page 6 13 System Verifications Page 6 14 Final Verificat...

Page 3: ......

Page 4: ...s illuminated when meter is charging Rear panel of the ILT5000 11 Power switch Turns meter off and on Turn meter ON up position prior to starting the software or connecting to a PC for the first time...

Page 5: ...vides an autorange mechanism allowing it to measure from 10 femptoamps to 1 milliamp of current While autoranging or using the various fixed current ranges settable via the various DataLight software...

Page 6: ...s Installer file with a msi extension When the msi file is executed you are brought through a simple set of Next Next Finish steps At the completion a DataLight folder will be placed on the desk top w...

Page 7: ......

Page 8: ......

Page 9: ...the Detector Current when setcurrentloop command was sent 4 20mA Current is linear in between Manual Current Loop Setting setcurrentloop 0 24 No relationship 4 20mA Current is set to whatever the setc...

Page 10: ...tically determine the appropriate security protocol WEP WPA and WPA2 are supported Set the IP Address This can either be accomplished by a Using DHCP which will automatically fill in the IP Address Ne...

Page 11: ......

Page 12: ......

Page 13: ...pply to damage resulting from improper set up accident alteration abuse loss of parts or repair by other than International Light Technologies The equipment will be repaired or replaced at our option...

Reviews: