4.  Zero the indicator and place calibrated test weights over each load cell in turn. 

The amount of test weights to be used will depend on the scale configuration; for 
specific recommendations, refer to Handbook 44 Field Manual, published by the 
Institute for Weights and Measures. For a four-

cell platform, it’s recommended 

using 25% of scale capacity. 

5.  Record the value displayed on the indicator after the test weight is placed in turn 

on each corner (directly over the load cell) without allowing the weight to 
overhand the sides. Allow the scale to return to zero each time to check for 
friction or other mechanical problems. Select the load cell which has the lowest 
value as your reference point. This cell will not be trimmed. 

6.  Replace the same test load over each cell in turn. Using the corresponding 

potentiometer, trim each cell down to equal the reference load cell. As corner 
corrections are somewhat interactive, check all cells again for repeatability. If 
necessary, repeat steps 4 and 5. 

7.  Pull excess cable out of the enclosure and tighten the cord grip assemblies with 

a wrench. To be watertight, each cord grip must be tightened so the rubber 
sleeve begins to protrude from the hub. 

8.  Unused hubs must be properly plugged to prevent moisture entry. If extra hole 

plugs are needed please consult factory. 

9.  Remove the desiccant from the plastic bag, and insert the desiccant bag into the 

junction box before closing. Inspect the desiccant during normal service and 
change desiccant as needed. 

10.  Replace the cover and tighten the cover screws in an alternating pattern to be 

certain the gasket is compressed equal in all locations. 

Summary of Contents for JB104SS

Page 1: ...rement INTERFACE 7401 E Butherus Drive Scottsdale AZ 85260 PHONE 480 948 5555 FAX 480 948 1924 www interfaceforce com Email gen interfaceforce com ORDER TOLL FREE 1 800 947 5598 JunctionBox Rev B 05 2...

Page 2: ...convenient for servicing and away from standing water Try to mount the enclosure in a location so that the load cell cables need not be cut nor length added Load cell output is temperature compensated...

Page 3: ...mblies and leave the grips loose until final closure Before connecting load cell wires to the terminals strip the wire insulation back 1 4 to expose the wire The spring loaded terminals will accommoda...

Page 4: ...924 www interfaceforce com Email gen interfaceforce com ORDER TOLL FREE 1 800 947 5598 JunctionBox Rev B 05 21 18 Figure 3 Application Use JB104SS Wiring Load Cell Color EXC RED EXC BLK SIG GRN SIG WH...

Page 5: ...ment INTERFACE 7401 E Butherus Drive Scottsdale AZ 85260 PHONE 480 948 5555 FAX 480 948 1924 www interfaceforce com Email gen interfaceforce com ORDER TOLL FREE 1 800 947 5598 JunctionBox Rev B 05 21...

Page 6: ...ERFACE 7401 E Butherus Drive Scottsdale AZ 85260 PHONE 480 948 5555 FAX 480 948 1924 www interfaceforce com Email gen interfaceforce com ORDER TOLL FREE 1 800 947 5598 JunctionBox Rev B 05 21 18 Figur...

Page 7: ...roblems The best trim is always the least amount of trim When all errors except the cell mismatch and cable extensions or reductions have been corrected continue with the trimming 3 1 Excitation Board...

Page 8: ...er mechanical problems Select the load cell which has the lowest value as your reference point This cell will not be trimmed 6 Replace the same test load over each cell in turn Using the corresponding...

Reviews: