1.1 

Introduction 

This instruction manual refers to the Interface model ITL tension link telemetry options. Before installing or 

operating any Interface tension link, this and any reference documents should be read and understood.

In compliance with the Machinery Directive, Interface tension link load cells will be supplied with the following 

documents: Instruction Manual and Calibration Certificate. 

1.2   Markings CE

Each tension link will be marked with an individual serial number, CE label and the SWL (safe working load) of 

the tension link.

1.3   Electromagnetic Compatibility (EMC)

The electromagnetic compatibility of the load cell device can only be assessed in conjunction with the entire 

installation, including its control systems. The machine builder who installs this partly completed machinery 

into a machine is responsible for compliance with the EMC directive. 

1.4 

Installation and Operation 

To ensure safe and trouble-free installation of the tension link measuring device, the tension link must be 

properly transported and stored, professionally installed and placed in operation. 

Unpacking

Before removing the tension link, inspect the packaging for signs of damage and immediately inform the 

supplier if any damage is found. Unpack the tension link carefully, taking special care with cables and be 

aware to the possibility of damaging low-range devices by mishandling. Ensure that calibration and instruction 

information is not inadvertently discarded. 

Checks Prior to Installation

• 

Check that the tension link has been suitably selected for the environment it is being placed into. Any 

adverse temperature, corrosive or potentially explosive environments may affect the operational life and 

the safety of the product. 

OPERATING INSTRUCTIONS

Installation

Tension link load cells are normally classified as portable devices, and so the correct installation is critical to 

ensuring product life cycle. To avoid damage or loss of accuracy during installation, the following points should 

be followed. 

• 

The direction of the load applied to a link should be linear as shown in Figure 1. 

Summary of Contents for ITL

Page 1: ...Helm Drive Scottsdale Arizona 85260 480 948 5555 www interfaceforce com Operator Instructions ISTL Tension Link v1 2 12 01 2017 Page 1 of 2 OPERATOR INSTRUCTIONS FOR MODEL ITL TENSION LINK LOAD CELL...

Page 2: ...ions ISTL Tension Link v1 2 12 01 2017 Page 2 of 2 TABLE OF CONTENTS 1 OPERATING INSTRUCTIONS 1 1 Introduction 1 1 2 Markings CE 1 1 3 Electromagnetic Compatibility EMC 1 1 4 Installation and Operatio...

Page 3: ...e 1 4 Installation and Operation To ensure safe and trouble free installation of the tension link measuring device the tension link must be properly transported and stored professionally installed and...

Page 4: ...om high power cables and equipment high output RF equipment and inductive loads and generators Tension Link Output When setting up your tension link the following points should be acknowledged The zer...

Page 5: ...ch the loading conditions that would be experienced in service but it is not possible to totally simulate the on site structure for every tension link manufactured It is for this reason that for optim...