background image

7

Integra

®  

Luxtec

® 

UltraLite® Plus Headlight

Part Number 

Description

AX1375BIF  

Headband with Ultralite 2 bifurcated cable, Ultralite module and extended linkage

AX1385BIF  

Sweatband with Ultralite 2 bifurcated cable, Ultralite module and extended linkage

 

AX1375BIFSL  

Headband with Ultralite 2 bifurcated cable and Ultralite module

AX1385BIFSL  

Sweatband with Ultralite 2 bifurcated cable and Ultralite module

 

001170BIF  

Replacement headband for AX1175SBIF

001190EL  

Module (gray) with extended linkage

001270BIF  

Replacement headband for AX1375BIF or AX1375BIFSL

 

001290BIF  

Replacement sweatband for AX1375BIF or AX1375BIFSL

 

400282

LX

Replacement kit pad for headband

0

o

0286  

Joystick for 001190, pack of 3

 
 

Replacement Parts for Ultralite® Plus Headlight 

Care and Maintenance

All Headlight Systems:

To ensure long life for the headband, cable, and modules, 

they should be stored in the duffel bag or case provided. Fiber 

optic cables are fine, high quality optical instruments. You can 

prolong the useful life by following a few guidelines:

•  Avoid stretching your cable, forming configurations  

  involving sharp angles or kinks, or contact with sharp or  

  pointed objects. The internal light fibers are made of glass,  

  a material which breaks under stress. Fiber breakage will

  result in diminished light output.

•  Utilize the gown clips (If provided) to keep the  

 

  headlight cable secured to the surgeon’s gown for greater  

  comfort and maneuverability. (This does not include  

  endoscopic use)

•  Do not use your fiber optic cable with any alterations to it’s  

  original design or fabrication. Bundle size (aperture) of the  

  fiber optic cable should be matched to the aperture of the

  instrument to obtain maximum light output.

•  NOTE: A larger aperture fiber optic cable will not increase  

  light output of a smaller aperture instrument.

•  Keep the optical faces from contacting the floor or other  

  hard surfaces. The resulting scratches will diminish light  

  output. Store the cables in sterilization trays for  

  additional protection.

•  Any inadvertent cut or puncture to the silicone tube will  

  render the cable unsafe. It should be taken out of  

  service immediately.

Cleaning Modules:

The headlight module may be wiped down with alcohol. The 

lens should be cleaned only with lens tissue available at any 

camera store. Follow the package directions.

Follow all applicable bloodborne pathogen cleaning 

procedures as indicated by OSHA and / or your

hospital requirements when cleaning, disinfecting,

and sterilizing instruments and accessories.

Cleaning Joystick (only):

The joystick can be removed from the module for autoclaving.

Cleaning Sweatband:

The sweatband may be cleaned using standard wash

cycles in warm water and detergent. Air dry.

Cleaning Headband:

The headband may be cleaned by an alcohol wipedown or  

washed with lukewarm water and mild detergent. Rinse  

thoroughly with water.

Cleaning Cable:

Wipe the fiber optic cable sheathing with a water dampened paper 

towel, cloth, or gauze before using 70 % Alcohol as the disinfectant.

Disinfecting Cable:

Soaking the UltraLite 2 cables in any sterilizing disinfectant  

should be avoided. There is a lens assembly at the light source  

end, which is not sealed and could allow moisture to penetrate 

behind the lensing.

Sterilizing Cable:

The UltraLite fiber optic cable has not been specifically  

designed to withstand sterilization. There is a lens assembly  

at the light source end, which is not sealed and could allow 

moisture to penetrate behind the lensing.

 

WARNING: Do not autoclave the headband,

 

cables, sweatband or headlight modules.

Summary of Contents for Luxtec UltraLite Plus

Page 1: ...OPERATION AND SERVICE MANUAL Integra Luxtec UltraLite Plus Headlight and Sweatband For use with high power surgical light sources EN Integra Luxtec UltraLite Plus Headlight and Sweatband...

Page 2: ...dlight Table of Contents Indication for Use 2 General Warnings 4 Assembly 5 6 Adjustments 6 Replacement Part Numbers 7 Care and Maintenance 7 Troubleshooting 8 Integra Equipment Warranty 8 Integra Cab...

Page 3: ...vasive While the cable light source endfittings are compatible with all light sources brightness and spot quality vary with specific types of light sources The UltraLite systems perform best with Inte...

Page 4: ...er Batch Code Serial Number Authorized Representative in the European Community Manufacturer Date of Manufacture This device meets CE requirements Caution High Voltage Equipotentiality Protective Eart...

Page 5: ...should be followed with special attention given to warnings controls and specifications Follow the manufacturer s instructions for other devices not recommended by this manual when used in conjuction...

Page 6: ...he fiber optic cable insert the angled end of the cable into the headlight module until it snaps into place Figure 2 A bifurcated cable fits into the cable clips on both sides and the rear of the head...

Page 7: ...stem by grasping the joystick The joystick can be removed for autoclaving or to prevent misaligning the headlight module when working in close quarters The iris adjustment varies the spot size from 20...

Page 8: ...obtain maximum light output NOTE A larger aperture fiber optic cable will not increase light output of a smaller aperture instrument Keep the optical faces from contacting the floor or other hard surf...

Page 9: ...device must be clean when returned to Integra Integra reserves the right to return unrepaired any equipment that is contaminated with blood or other organic material To obtain service under warranty o...

Page 10: ...ation or its subsidiaries in the United States and or other countries 2019 Integra LifeSciences Corporation All rights reserved 60905994 Rev A 1430078 1 Integra Luxtec UltraLite Plus Headlight www int...

Reviews: