Instruction Manual VGC301 Controller
InstruTech, Inc.
Page 24
7.2
Log 1-8; Log-Linear Analog Output Equation & Table - mbar
Log-Linear 1 to 8 V analog output for selected gases - Engineering units in mbar
True
Pressure
(mbar)
N
2
Ar
He
O
2
CO
2
KR
Freon12 Freon22
D
2
Ne
CH
4
0.0001
1.000
1.000
1.000
1.000
1.000
1.000
1.000
1.000
1.000
1.000
1.000
0.0002
1.301
1.301
1.301
1.301
1.301
1.301
1.301
1.301
1.301
1.301
1.301
0.0005
1.699
1.699
1.699
1.699
1.699
1.523
1.699
1.699
1.699
1.699
1.699
0.0010
2.000
1.903
1.938
2.000
2.028
1.668
2.125
2.125
2.080
1.903
2.167
0.0020
2.301
2.146
2.204
2.301
2.355
1.970
2.487
2.487
2.392
2.166
2.523
0.0050
2.699
2.524
2.602
2.699
2.672
2.370
2.883
2.855
2.778
2.551
2.893
0.0100
3.000
2.820
2.908
2.991
3.012
2.675
3.172
3.136
3.082
2.849
3.186
0.0200
3.301
3.188
3.208
3.294
3.345
2.979
3.473
3.434
3.385
3.150
3.484
0.0500
3.699
3.512
3.607
3.693
3.741
3.372
3.863
3.837
3.779
3.543
3.886
0.1000
4.000
3.809
3.928
3.989
4.033
3.671
4.157
4.136
4.082
3.844
4.197
0.2000
4.301
4.103
4.217
4.288
4.325
3.963
4.445
4.424
4.393
4.148
4.500
0.5000
4.699
4.495
4.634
4.686
4.696
4.341
4.798
4.783
4.828
4.553
4.893
1.0000
5.000
4.784
4.962
4.987
4.982
4.614
5.044
5.037
5.174
4.867
5.201
2.0000
5.301
5.064
5.324
5.288
5.249
4.865
5.250
5.255
5.579
5.192
5.517
5.0000
5.699
5.404
6.070
5.695
5.550
5.141
5.447
5.471
7.288
5.696
5.877
10.0000
6.000
5.633
8.125
6.008
5.743
5.309
5.556
5.602
8.125
6.252
6.374
20.0000
6.301
5.815
8.125
6.337
5.886
5.433
5.621
5.675
8.125
7.608
7.409
50.0000
6.699
5.969
8.125
6.862
6.002
5.514
5.680
5.722
8.125
8.125
7.930
100.0000
7.000
6.045
8.125
7.282
6.065
5.548
5.751
5.780
8.125
8.125
8.125
200.0000
7.301
6.093
8.125
7.526
6.157
5.606
5.851
5.877
8.125
8.125
8.125
300.0000
7.477
6.131
8.125
7.625
6.253
5.654
5.918
5.950
8.125
8.125
8.125
400.0000
7.602
6.178
8.125
7.705
6.353
5.679
5.962
6.000
8.125
8.125
8.125
500.0000
7.699
6.237
8.125
7.786
6.448
5.710
5.996
6.038
8.125
8.125
8.125
600.0000
7.778
6.295
8.125
7.861
6.532
5.734
6.025
6.070
8.125
8.125
8.125
700.0000
7.845
6.349
8.125
7.928
6.611
5.754
6.050
6.097
8.125
8.125
8.125
760.0000
7.881
6.380
8.125
7.965
6.658
5.765
6.063
6.112
8.125
8.125
8.125
800.0000
7.903
6.399
8.125
7.988
6.687
5.772
6.072
6.122
8.125
8.125
8.125
900.0000
7.954
6.488
8.125
8.042
6.766
5.787
6.092
6.146
8.125
8.125
8.125
1000.0000
8.000
6.494
8.125
8.092
6.847
5.799
6.111
6.167
8.125
8.125
8.125
1100.0000
8.041
6.539
8.125
8.125
6.936
5.812
6.128
6.187
8.125
8.125
8.125
1200.0000
8.079
6.580
8.125
8.125
7.028
5.822
6.146
6.204
8.125
8.125
8.125
1300.0000
8.114
6.624
8.125
8.125
7.140
5.828
6.164
6.222
8.125
8.125
8.125
1333.0000
8.125
6.636
8.125
8.125
7.169
5.830
6.169
6.228
8.125
8.125
8.125
Values listed under each gas type are in volts.
The log-linear output signal and pressure are related by the following formulas:
P = 10
(V - 5)
V = log
10
(P) + 5
where P is the pressure in mbar, and V is the output signal in volts.
An analog output of 10 volts indicates a faulty convection gauge or unplugged gauge cable.
The chart on the following page shows the graphical results of the table and formulas given above for nitrogen.
True pressure (N
2
) is plotted on the X-axis with a log scale. The output signal is plotted on the Y-axis on a linear scale.
Note - when using the units of pascals and LOG 1-8 is selected , the same equation of P = 10
(V - 5)
listed above
applies. This results in a log-linear analog output range of 1 to 10.12 Vdc for 1.30E-02 pascals to 133 kpa.