![Instrutech The Hornet IGM402 User Manual Download Page 8](http://html1.mh-extra.com/html/instrutech/the-hornet-igm402/the-hornet-igm402_user-manual_2070238008.webp)
Instruction Manual IGM402 Hornet
InstruTech
Page 6
Specifications - continued
setpoint relays (3 total)
three single-pole, double-throw (SPDT), 1 A at 30 Vdc resistive, or ac non-inductive
status outputs
degas & filament on/off status via display messages, open collector transistor or RS485
input signal
degas and filament on/off & emission current are set by continuity to ground
using digital inputs, RS485 or manually using front panel soft-keys
filament selection
filament 1 or 2 selectable via front panel soft-keys or RS485 commands
input power
20 to 28 Vdc, 30 W protected against power reversal and transient over-voltages
connectors
(2) 9-pin D-Sub, (2) terminal blocks, (2) convection gauge connectors
convection gauge compatibility
InstruTech CVG101
Worker Bee™
or Granville-Phillips
®
275 Convectron
®
CE compliance
EMC Directive 2014/30/EU, EN55011, EN61000-6-2, EN61000-6-4, EN61326-1, EN61010-1
environmental
RoHS
1.3
Dimensions
fitting dimension A
NW16KF
1.45 in. (37mm)
NW25KF 1.45 in. (37mm)
NW40KF 1.45 in. (37mm)
1 1/3 in. Mini CF 1.85 in. (47 mm)
2 3/4 in. Conflat® 1.70 in. (43 mm)
3/4 in. Tube
2.16 in. (55 mm)
1/2 in. VCR
2.58 in. (65 mm)
1.4
Part Numbers
IGM402 Fittings / Flanges
With Yttria
Filaments
With Tungsten
Filaments
Replacement
Sensor - Yttria
Replacement
Sensor - Tungsten
NW16KF
IGM402YBD
IGM402TBD
IG4YB
IG4TB
NW25KF
IGM402YCD
IGM402TCD
IG4YC
IG4TC
NW40KF
IGM402YDD
IGM402TDD
IG4YD
IG4TD
1 1/3 in. Mini-CF/NW16CF Mini-Conflat®
IGM402YED
IGM402TED
IG4YE
IG4TE
2 3/4 in. CF / NW35CF Conflat® IGM402YFD
IGM402TFD
IG4YF
IG4TF
3/4 in. Tube (3/4 in. O.D. O-ring compression)
IGM402YAD
IGM402TAD
IG4YA
IG4TA
1/2 in. Cajon® 8VCR® female
IGM402YHD
IGM402THD
IG4YH
IG4TH
IGM402 / CVG101 convection gauge cable
P/N
Cable, IGM402 to CVG101 , 3 ft. (1m)
HB431-1-03F
Cable, IGM402 to CVG101, 10 ft. (3m)
HB431-1-10F
Cable, IGM402 to CVG101, 25 ft. (8m)
HB431-1-25F
Cable, IGM402 to CVG101, 50 ft. (15m)
HB431-1-50F
Note -
For convection gauge product and installation information refer to CVG101 Worker Bee™ User Manual.