Chapter: Requirements
30
M10-17417-EN
Noise levels do not include noise from specimen breaks.
The noise values are measured at the ear level of an operator positioned in front of the
machine.
The noise level of the entire system depends on your particular system configuration.
Refer to the documentation supplied with other system components for information
about that component's noise level.
The noise output from equipment used for materials or structures testing is also depen-
dent upon the items under test. Instron recommends that users carry out their own noise
level measurements to ensure the continuous safety and comfort of personnel.
68TM-10 Power Requirements and Cords
Power requirements
For 230 Vac operation, the frame is normally set to 240 V.
Warning
Electrical hazard - to ensure safety and Electromagnetic Compatibility (EMC) the
facility’s receptacle must be a 3-wire grounded receptacle. The ground must be a
low impedance earth ground in accordance with national and/or local regulations.
Table 5. 68TM-10 power requirements
Parameter
Specification
Maximum Power (VA)
1400
Single Phase Voltage (Vac) (±10%)
100, 120, 220, 240
Frequency - Hz
47 to 63
Length of Power Cable - m (ft)
2.44 (8)
Summary of Contents for 68TM-10
Page 1: ...Model 68TM 10 Dual Column Table Model Preinstallation Manual M10 17417 EN Revision B ...
Page 8: ...Preliminary Pages 8 M10 17417 EN ...
Page 11: ...11 Preliminary Pages Product Support www instron com Crane method 61 Index 63 ...
Page 12: ...Preliminary Pages 12 M10 17417 EN ...
Page 52: ...Chapter Risk reduction and safe use 52 M10 17417 EN ...
Page 65: ......
Page 66: ...www instron com ...