![Instron 1000KPX-J3D Operating Instructions Manual Download Page 24](http://html1.mh-extra.com/html/instron/1000kpx-j3d/1000kpx-j3d_operating-instructions-manual_2070125024.webp)
Chapter 2: Specifications
24
M47-17035-EN
Hydraulic power supply physical dimensions
59 Series control unit
Table 6. Dimensions of standard HPS models.
Height:
1038 mm (40.9 in)
Required Floor Space (W x D)
1
:
1365 x 959 mm (53.75 x 37.75 in)
Weight:
680 kgs (1,500 lbs)
1. Width does not include clearance to open the electrical box door. Clearance needed to open the
door is 355 mm (14 in).
Table 7. 59 Series control unit technical specifications.
Data Sampling
40 kHz
Data Capture
Selectable up to 1000 Hz
1
Intelligent data capture
Synchronous on all channels
Digital Signal Processor
32-bit floating point
Self-test diagnostics
Real-time closed-loop control
Real-time data acquisition
Data Transfer to Computer
1000 Hz
Transducer Inputs
Standard Instron rationalized transducers
Any 0-10 V analog DC input
Transducer Resolution
1 part in 500,000 of +/- full scale (19 bits)
Load Measurement Accuracy
+/- 0.5% of reading down to 1/500 of load cell capacity
Linearity
+/- 0.25% of reading over a range of 0.2% to 100% of capacity
Repeatability
+/- 0.25% of reading over a range of 0.2% to 100% of capacity
Strain Measurement Accuracy
+/- 0.5% of reading down to 1/50 of full range with ASTM E 83 class B-1 or B-2,
or ISO 9513 class 0.5 extensometer
Certifications
Conform to all relevant European standards
1. Software data capture rate may vary.