![Instron 1000KPX-J3D Operating Instructions Manual Download Page 22](http://html1.mh-extra.com/html/instron/1000kpx-j3d/1000kpx-j3d_operating-instructions-manual_2070125022.webp)
Chapter 2: Specifications
22
M47-17035-EN
Overall Frame Area (width (F) x depth (G))
J1
1162 x 864 mm
(45.75 x 34 in)
1283 x 1219 mm
(50.5 x 48 in)
1473 x 1219 mm
(58 x 48 in)
N/A
J3
1568 x 864 mm
(61.75 x 34 in)
1687 x 1219 mm
(66.438 x 48 in)
1878 x 1219 mm
(73.938 x 48 in)
1816 x 1410 mm
(71.5 x 55.5 in)
J4
N/A
N/A
N/A
1331 x 1241 mm
(52.375 x 48.875 in)
Weight
7
Refer to the appropriate General Assembly drawing provided in the “Reference Manual-
Equipment”
Speed Information
Testing Speed Range
(at full load)
0.1 to 203 mm/min (0.004 to 8.0 in/min)
Jog (Adjusting) Speed
(Maximum)
203 mm/min (8.0 in/min)
Position Information
Accuracy
8, 9
C3
Equal to or less than +/-0.13 mm (0.005 in) or +/-0.1% of displayed reading, whichever is
greater
C4
Consult factory
Resolution
8, 10
C3
1.0 µm (0.00004 in)
C4
Consult factory
Cyclic Test and Frequency
Limits
Capable of tension/tension or compression/compression cyclic testing only. Cyclic tests are
limited to ramp type waveforms as defined within the materials testing software package being
used with the system. Frequency limits are defined by the compliance of the specimen and the
dynamic control limit of the system. In all cases, the frequency for cyclic testing is limited to
1 Hz or less.
Parallelism and Alignment
11
Parallelism
12
0.2 mm (0.008 in)
Center Alignment
J1, J3
0.5 mm (0.02 in)
J4
0.8 mm (0.03 in)
Power Requirements
Power option A
208/230 VAC, 3 Ph, 50/60 Hz, 30 A
208/230 VAC, 3 Ph,
50/60 Hz, 40 A
Power option B
380/400/415 VAC, 3 Ph, 50/60 Hz, 20 A
380/400/415 VAC,
3 Ph, 50/60 Hz, 30 A
Power option C
460 VAC, 3 Ph, 50/60 Hz, 15 A
460 VAC, 3 Ph,
50/60 Hz, 25 A
Certifications
Conform to all relevant European standards and carry a CE mark
Table 5. KPX frame technical specifications. (Continued)
Specification
1
600KPX
1000KPX
1500KPX
2000KPX