User manual |
Specifications and safety
42
Protection index:
IPX1.
Classification:
Class I, internally energized. CF Type.
Operating mode:
Continuous operation (frequent).
Screen:
Size: 99 mm x 19 mm.
Type: alphanumeric LCD.
Environmental specifications
Temperature:
Operational: 0 to 50ºC.
Storage: 0 to 70ºC.
Humidity:
Operational: 10 to 95% RH, without condensation
Storage: 10 to 95% RH, without condensation
WARNING: if the Apolus is used outside these conditions, 15 through 30
minutes will be required to stabilize the system so that functioning failures
do not occur.
Defibrillator
Waveform:
Biphasic truncated exponential. Wave shaped parameters
adjusted according to the patient’s impedance.
Shock application:
By means of multifunction pads or external adult/child pads.
Scales for adult/external
defibrillation:
5, 10, 15, 20, 30, 40, 50, 80, 100, 150, 200, 250, 300 and
360 Joules. Maximum energy limited to 50 Joules in internal
or child pads.
Operation:
Standard sequence “1-2-3”
Selector key:
Allows turning the equipment on and off, selecting energy
scales and activating functional test mode.