![Instramed APOLUS User Manual Download Page 33](http://html.mh-extra.com/html/instramed/apolus/apolus_user-manual_2070094033.webp)
User manual |
Care and maintenance
33
Electromagnetic emissions
Directives and manufacturer declaration - electromagnetic emissions
The Apolus is intended for use in the specific electromagnetic environment below. The customer or user of the
defibrillator is advised to ensure that it is used in such an environment.
Tests
Compliance
Electromagnetic environment - directives
RF Emissions ABNT NBR
IEC CISPR11
Group 1
The Apolus only uses RF power for its internal
functions. Nevertheless, its RF emissions
are very low and are not likely to cause any
interference in nearby electronic equipment.
RF Emissions ABNT NBR
IEC CISPR11
Class B
The Apolus is suited for use in any
establishment. This includes residential
establishments and those directly connected
to the public network of distribution of low
voltage electricity which supply domestic use
buildings.
Harmonics emissions
IEC 1000-3-2
Class A
Voltage fluctuations/flicker emissions
IEC 61000-3-3
Complies
NOTE: It is essential that the actual effectiveness of the RF shielding and the actual RF filter attenuation of the
shielded location be checked to ensure that they meet or exceed the specified minimum values.