![insize IST-TT Series Operation Manual Download Page 2](http://html1.mh-extra.com/html/insize/ist-tt-series/ist-tt-series_operation-manual_2068579002.webp)
1
Attention
8
Communication
1
Connecting
communication cable
Turn
off
power
and
then
connect
the
accessory
cable between
the COM port of PC and torque tester.
Uploading record date
2
5
Change the tester operation mode to “ ”.(Please refer to “
Peak
Make sure the connection between PC and tester is normal.
In uploader program, first select the correct COM port No.
Next, select the file path to save the uploaded data.
Use PC to start the uploader program.
Send
Finally, press “upload” button to transmit the torque records to PC.
The uploaded date is then shown on the column and saved in the
Hold Mode Recorded Value Review
” section)
*
.csv file. Use Microsoft Excel to view
*
.csv file.
Keep work area clean.
Do not use the torque transducer with impact or impulse tools.
Ensure that components are properly connected before using torque
Over-torque (110% of Max. torque range) could cause breakage or
lose accuracy.
tester.
Do not apply excessive force to the LCD panel.
Do not attempt to disassemble or repair the torque tester. It may
Do not use organic solvents, such as alcohol or paint thinner when
cleaning the torque tester.
Do not use unpermitted sockets or accessories.
Do not permit unauthorized personnel to operate the torque tester.
damage torque tester.
Wear proper clothing to avoid loose articles being caught in moving
parts.
When not in use, torque tester should be stored in proper place.
electric shock.
Do not handle the power plug with wet hands. Avoid causing an
2. The “Peak hold” mode recorded value review also can be operated
in “Track” mode operation.
3. Communication mode is for uploading recorded data to PC.
8
Setting
tolerance
Long press key after turn on the instrument, the screen start
flickering.
Select the preset value number by press up and down keys,
press key enter next setting item.
M
U/S
Measured value is within the percentage range of target: Judge
OK!
Setting range can be 25.0% ~ -25.0%, ± 00.0% means no
judgement.
Judgement function is only effective in P / 1P mode.
U/S
Set the measuring mode, unit, preset value, up and
down
limit
percentages, press up and down
keys
to
adjust, press
the
key enter next setting item.