Specification:

3

Display range: 0 999HLD

~

Accuracy: ±6HLD(when HLD

=

800)

Conversion scales: HV, HB, HRB, HRC, HV, HSD, MPa

repeatability:  6HLD

±

Impact direction: 

0

°, 

°, 180°, 

°, 

°

±90

±45

±135

Button function:

4

LCD display:

5

Leeb Value Calibration

Please  do  calibration  when  use  the  device  first  time  or  interval  for  a 
long time.

press ruturn button and On/Off button together (about 2 seconds) 
to  enter  Leeb  value  calibration,  the  INSIZE  logo  will  blink  on  top 
left  corner.  Measure  5  times  on  the  surface  of  standard  test 
block(you should do enter calibration mode again if get any value 
which deviation is large).

2

Input  the  actual  HLD  value  of  the  test  block  by  use  up  and  down 
buttons, then press return button back to measurement mode.

3

Note: The distance between any two impact position should be 

3mm. 

the  distance  between  impact  position  and  the  edge  of  the  sample 
should be 

3mm.

1

3

5

4

2

 

 

1. On/Off button    : Turn it on/off by pressing and holding the 
button.
2. Up arrow button     : 

①  

In  measurement  mode,  press  this  button  to  select 

conversion scale.

②  

In  measurement  mode,

 

press 

   

button  to  delete  current 

result and press this button to confirm. 
3. Setup button     :

① 

In measurement mode, press this button to select the items 

need  to  be  set  like  statistic  times,  material  code,  date,  time, 
etc.

②  

Turn  off  the  machine,  press  this  button  and  On/Off  button 

together to enter conversion value calibration.
4. Return button    :

① 

Back to measurement mode.

②  

Turn  off  the  machine,  press  this  button  and  On/Off  button 

together to enter Leeb value calibration.
5.  Down  arrow  button      :  In  measurement  mode,  press  this 
button to enter data browsing mode.

 

 

 

 

Date and time

Material code

 

Conversion value

Impact direction

 

 Impact and statistic times 

HLD value

HLD  averaged value

Note: Please do calibration with standard test block in packing box. 

Press setup button then set measuring conditions by using up and 
down  buttons,  select  material  code  to  number  01,  the  impact 
direction is down. Turn off the machine after do setup.

1

Conversion Value Calibration

This  calibration  is  applied  to  some  workpiece  you  don’t  know  the 
Leeb hardeness but only conversion hardness value like HB, HV, etc.

Note: Please do calibration with known hardness workpiece.

Press up arrow button to select the conversion scale, then Press 
setup button then set measuring conditions by using up and down

1