3
4
Specification:
3
Display range: 0 999HLD
~
Accuracy: ±6HLD(when HLD
=
800)
Conversion scales: HV, HB, HRB, HRC, HV, HSD, MPa
repeatability: 6HLD
±
Impact direction:
0
°,
°, 180°,
°,
°
±90
±45
±135
Button function:
4
LCD display:
5
Leeb Value Calibration
Please do calibration when use the device first time or interval for a
long time.
press ruturn button and On/Off button together (about 2 seconds)
to enter Leeb value calibration, the INSIZE logo will blink on top
left corner. Measure 5 times on the surface of standard test
block(you should do enter calibration mode again if get any value
which deviation is large).
2
Input the actual HLD value of the test block by use up and down
buttons, then press return button back to measurement mode.
3
Note: The distance between any two impact position should be
≥
3mm.
the distance between impact position and the edge of the sample
should be
≥
3mm.
1
3
5
4
2
1. On/Off button : Turn it on/off by pressing and holding the
button.
2. Up arrow button :
①
In measurement mode, press this button to select
conversion scale.
②
In measurement mode,
press
button to delete current
result and press this button to confirm.
3. Setup button :
①
In measurement mode, press this button to select the items
need to be set like statistic times, material code, date, time,
etc.
②
Turn off the machine, press this button and On/Off button
together to enter conversion value calibration.
4. Return button :
①
Back to measurement mode.
②
Turn off the machine, press this button and On/Off button
together to enter Leeb value calibration.
5. Down arrow button : In measurement mode, press this
button to enter data browsing mode.
Date and time
Material code
Conversion value
Impact direction
Impact and statistic times
HLD value
HLD averaged value
Note: Please do calibration with standard test block in packing box.
Press setup button then set measuring conditions by using up and
down buttons, select material code to number 01, the impact
direction is down. Turn off the machine after do setup.
1
Conversion Value Calibration
This calibration is applied to some workpiece you don’t know the
Leeb hardeness but only conversion hardness value like HB, HV, etc.
Note: Please do calibration with known hardness workpiece.
Press up arrow button to select the conversion scale, then Press
setup button then set measuring conditions by using up and down
1