insize ISH-PHB Operation Manual Download Page 5

7

8

larger than 30mm. And appropriate supporting ring should be 
elected and installed.
7)  A  sample  should  not  be  with  magnetic.  The  signal  of  the 
impact device would be seriously interfered in the work by the 
magnetic, which may cause inaccurate test results. 

2

Measuring Steps:

Loading: Simply load the impact device by sliding the loading 
tube forward.

Place: Then  place  and  hold  the  impact  device  on  the  surface 
of  the  test  piece  at  the  desired  test  point.  Impact  indirection 
should be vertical with the test of surface.

Burst  impact  (Measure):  Trigger  the  impact  by  pressing  the 
release  button.  The  hardness  value  will  be  instantaneously 
displayed.

Read off the test result from screen.

Note:  Generally,  each  measurement  location  of  sample  is 
conducted  for  the  five  tests.  The  “S”  (difference  of  maximum 
value  and  minimum  value)  values  must  be  less  than  15HL. 
The  distance  between  any  two  impact  position  should  be  ≥3 
mm ; the distance between impact position and the edge of the 
sample should be ≥3 mm.

Memory:

3

The  data  group  (such  as  test  result,  conversion  result,  sample 
material and impact direction) automatically save in memory after 
one individual measurement. ISH-PHB hardness tester can store 
99  sets  of  data,  when  measuring  times  more  than  99,  the  last 
group  of  data  will  store  the  99th  position  and  the  first  group  of 
data will be erased. The second group of data will be moved into 
the  2th  position,  simultaneously  the  position  of  other  group  of 
data, will be moved into lower position.
Press down arrow button to enter memory store mode, use up and 
down buttons to browsing the memory.
Print (the printer is optional):

4

After  one  set  measurement  press  setup  button  and  let  the  print 
symbol blink, now press return button to print current result.

After  using  1000-2000  times,  users  should  clean  the  catheter  of 
impact  device  and  impact  body  with  nylon  brush,  and  screw  off 
the  supporting  ring  before  clean  the  catheter,  and  then  take  out 
the  impact  body,  rotate  the  nylon  brush  into  the  tube  in  an  anti-
clockwise  direction,  and  pull  out  when  touching  the  bottom.  So 
repeatedly, and then load up the impact body and supporting ring; 
Users should release the impact body after use. And the lubricant 
is banned.

In calibration of the hardness tester, if finding that error is larger 
than  6HLD,  users  must  be  renew  the  steel  ball  or  impact  body, 
because  the  reason  may  be  that  the  steel  ball  or  impact  body  is 
wore out to lead to failure in operation.

Maintenance

1

2

Impact Device Maintenance:

Normal Maintenance Procedures: