insize ISH-MRD200 Operation Manual Download Page 4

5

6

scale

indenter

test force(N)

HRA

HRB

HRC

diamond taper indenter

(cone angle

)

120°

ball indenter

(ball dia.=1.5875mm)

diamond taper indenter

(cone angle120°)

588

980

1471

HRD

HRF

HRG

diamond taper indenter

(cone angle120°)

ball indenter

(ball dia.=1.5875mm)

588

1471

980

ball indenter

(ball dia.=1.5875mm)

f. The measuring surface of the test material should be clear 
and without greasy dirt, oxide skin, grooves or clear 
machining traces, and the roughness of the measuring 
surface no more than 0.8µm is proposed. 

 

scale, test force and weight application form

 

test force(N)

588
980

1471

 

scale

HRA/HRF
HRB/HRD

HRC/HRG

a

fter installation, put the up cover and back cover on.

weight

d weight A

d weight A + weight B

d weight A + weight B + weight C

Usage

3.1 Preparation before use

      a. Choose the right stage according to the shape and size of 
the test sample.

      b. Choose indenters and forces according to the material and 
hardness scale(as below table shows).

Material(example)

carbide

 

steel

surface

carburizing and 

quenching steel

mild steel, aluminum alloy,

copper alloy, malleable

iron, annealed steel

annealed steel,

tempering steel,

alloy steel, stainless steel

sheet steel,

case-hardening steel

magal, mild cast iron,

thin mild steel plate,

annealed copper alloy

phosphor bronze,

beryllium bronze,

malleable iron

c. Install indenters: put the indenter to spindle hole, ensure 
the spindle hole, end face, end handle of indenter and shaft 
shoulder clear with nothing. Tighten the lock screw at the end 
handle of indenter.

Attention

: take the indenter tip with fingertip when 

installation to avoid the indenter falling down and cause 
indenter tip damaged.

d. Rotate the test force dial to choose test force.

Attention: 

choose test force should go on under unload state.

e.  Adjust  load  dwell  time  to  five  seconds  from  adjusting  “+” 
and “-”buttons.

Reviews: