If the self test fails, the user is advised to check the following:
•
No Oxygen Mixer connected to the ventilator
•
No oxygen reservoir
•
Dirty/obstructed inlet filter
•
Pressure and control tube connected properly (kinks, disconnec-
tion or rupture of the tubes)
•
No HME or filter are connected to the exhalation valve
•
Patient is not connected
9.3. PATIENT WEIGHT
Rotate the scroll knob to select the weight of the patient:
The patient circuit type will be identified automatically, and the appropriate
weight screen (Pediatric/Adult) will be displayed.
Pediatric and adult weight selection screens
Note:
Setting patient weight will automatically set all ventilation
parameters.