14
5.
Remove High and Low Speed
Handpieces again.
6.
Place four drops of
Bio Lube™
oil (pen)
into the air line (See Figure 6.3).
Reattach, and then depress the foot
pedal blowing air through the
handpiece for 15 to 20 seconds. (See
Figure 6.5)
•
Inovadent recommends using Bio
Lube™ Cleaner for the cleaning and
lubricating process. (See Figure 6.9)
7.
Switch the water bottles switch to the OFF
position (See Figure 6.6).
8.
If chlorhexidine was used in one bottle,
remove that water bottle, rinse and allow
the bottle to dry.
Note:
If using a chlorhexidine solution, the
solution must be purged from the system
using water.
9.
Remove KLAW
™
tip from Piezoelectric
Handpiece, carefully clean and place in
storage box
.
10.
Remove bur from High Speed Handpiece
and replace with blank bur
.
11.
Store handpieces in original storage
containers (High, Low and Piezoelectric).
12.
Unplug cart from wall outlet if storing cart
in another location or turn off power strip
.
13.
Wipe down cart, cords and handpieces with
water and non-chloride non-iodine
disinfectant
.
14.
Replace all handpiece cords to rack to keep
cords clean and from being stepped on
(See Figure 6.7).
15.
Place the tank drainage hose end into small
bottle and crack open drain valve slightly to
expel moisture from the tank. Close drain
valve after all moisture has been expelled.
(See Figure 6.8).
Figure 6.6
Figure 6.5
Figure 6.7
Figure 6.8